Welcome to the Industrial Automation website!

NameDescriptionContent
XING-Automation
E-mail  
Password  
  
Forgot password?
  Register
当前位置:

Sysmex XN-L blood analyzer maintenance and troubleshooting

F: | Au:FANS | DA:2026-09-23 | 20 Br: | 🔊 点击朗读正文 ❚❚ | Share:

In terms of precision, the repeatability CV of WBC in whole blood mode is not greater than 3.0%, RBC is not greater than 1.5%, HGB is not greater than 1.5%, HCT is not greater than 1.5%, and PLT is not greater than 4.0%. In pre dilution mode, the corresponding CV will be relaxed, such as WBC not exceeding 5.0%, RBC not exceeding 4.5%, and PLT not exceeding 12.0%. In terms of carrying pollution, WBC, RBC, HGB, HCT, PLT usually require no more than 1.0%, and white blood cell classification items generally do not exceed 2.0% or specific absolute values. The stability of the sample after collection is affected by temperature. WBC, RBC, and HGB can remain stable for about 72 hours at 18 to 26 ℃, HCT for about 8 hours, and PLT for about 48 hours; Under the conditions of 2 to 8 ℃, some parameters have different stabilization times. Before testing, it is necessary to restore room temperature and mix thoroughly.

In terms of software data storage capacity, it can store approximately 100000 samples, 10000 patient information, 200 wards, 200 doctor names, 2000 analysis registration records, 20 calibration logs for each analyzer, 99 QC files for each analyzer, 300 image points per file, 5000 reagent replacement logs, and 5000 maintenance logs. Quality control supports L-J control chart, approximately 300 points x 96 files, and X-barM control chart, 300 points x 3 files.

Reagents, consumables, and calibration systems

The XN-L series uses a designated reagent system. CELLPACK DCL is a diluent and sheath containing sodium chloride, Tris buffer EDTA-2K, Store at a temperature of 2 to 35 ℃ for 60 days after opening the bottle. CELLPACK DST is a concentrated reagent that needs to be connected to the designated reagent preparation device of Sysmex, stored at 15 to 30 ℃, and opened for 60 days. SULFULYSER is used for hemoglobin measurement, containing sodium dodecyl sulfate, stored at 1 to 30 ℃, and opened for 60 days. Lysercell WDF is used for hemolysis and leukocyte treatment, 90 days after opening the bottle. Fluorocell WDF is a fluorescent dye containing polyacetylene pigment, methanol, and ethylene glycol. It should be stored in the dark at 2 to 35 ℃ and opened for 90 days. CELLCLEAN AUTO is a strong alkaline cleaning solution containing sodium hypochlorite, used for pipeline cleaning and stored at 1 to 30 ℃.

Both the quality control material XN-L CHECK and the calibration material XN CAL need to be refrigerated at 2 to 8 ℃ in the dark. Use within 15 days after opening XN-L CHECK and within 4 hours after opening XN CAL. The calibration items cover WBC, RBC, HGB, HCT, and PLT. The quality control barcode usually uses CODE128, which is a fixed string of "QC -" followed by an 8-digit batch number and a 1-digit checksum. If barcode reading fails, the system may automatically assign a sample number starting with ERR; QC samples, background checks, precision checks, and calibration samples also have specific prefixes. Daily use should avoid using expired reagents, repeatedly connecting opened bottles of reagents, and using non specified sample tubes or sampling adapters. A specialized adapter is required for the 15 mm sample tube. The metal parts of the RBT micro collection tube and the conventional sample tube on the sampler are positioned differently, and incorrect installation may result in sampling failure or sample drop.


Common faults and troubleshooting strategies

1. Sampling and barcode anomalies. If barcode reading fails, first check the label position. The XN-530 barcode label should be affixed within the designated area, and the barcode strip should be horizontal without excessive attachment, wrinkling, exceeding the bottom of the tube, or using a reusable cap. Check if the barcode type is within the supported range, such as ITF, CODABAR/NW7, CODE39, JAN/EAN/UPC, ISBT128, CODE128. If using a checksum, it should be confirmed that the setting is correct. The metal parts of the sampling adapter must be installed according to the type of sample tube and cannot be disassembled or reinstalled at will, otherwise it may become loose due to the expansion of screw holes.

2. Hole blockage, bubbles, and sampling noise. In the service data, the RBC/PLT page can view sampling data, particle counts of RBC and PLT, differences between maximum and minimum values, Clog and Bubble values. If the hole is blocked or the bubble value is abnormal, the cleaning procedure should be performed, using CELLCLEAN AUTO to clean the pipeline. Avoid generating bubbles during reagent connection, as bubbles can cause abnormal analysis. If the reagent is contaminated, turbid, or discolored, it should be replaced. When maintaining areas such as the RBC detector cover, air pump unit, and pneumatic capture chamber, power should be turned off and attention should be paid to the high temperature on the surface of the air pump.

3. Abnormal results and interference. Pseudo low WBC can be seen in white blood cell aggregation; Pseudo elevated levels can be seen in platelet aggregation, hemolysis, nucleated red blood cells, red blood cell aggregation, cold agglutinin, chyle, cold protein, cold globulin, fibrin, giant platelets, etc. Pseudo low RBC can be seen in red blood cell aggregation, small red blood cells, and red blood cell debris; Pseudo elevated levels can be seen in leukocytosis greater than 100000/μ L and giant platelets. Pseudo elevated HGB can be seen in significantly increased white blood cells, lipid levels, and abnormal proteins. Pseudo low HCT can be seen in cold agglutination, small red blood cells, and red blood cell fragments; Pseudo high can be seen in leukocytosis, severe diabetes, uremia, spherocytosis. Pseudo low PLT can be seen in platelet aggregation, pseudo thrombocytopenia, and giant platelets; Pseudo elevated levels can be seen in small red blood cells, red blood cell fragments, white blood cell fragments, cold proteins, and cold globulins. Research on interfering substances shows that bilirubin C, bilirubin F, hemolysis, fat emulsion, and chyle have concentration thresholds that affect WBC, RBC, HGB, HCT, and PLT. If the threshold is exceeded, they should be treated as interference.

  • Sigmatek TAE732-P 01-240-732-P Touch Display Unit
  • Sigmatek MDM021 Digital Mixed Module
  • Sigmatek DCP642 DIAS Central Unit
  • Sigmatek CST022 12-014-022 C-DIAS Module
  • Sigmatek SRO021 20-893-021 Safety Relay Output
  • Krauss Maffei MC5 Control System Sigmatek
  • Sigmatek DC061 Module 18-24VDC 300mA
  • Herrmann EVT0855 Ultrasonic Welder Touch Panel
  • Sigmatek CIV521 Control Module 1.4A 2W
  • Sigmatek DDI61 05-006-161 Digital Input Module
  • SIGMATEK ETT312-E Touch Terminal 01-230-312
  • Cessna S3326-1 Sigma Tek 5000B-67 Gyro 1U149-015-9
  • SIGMATEK PS101 Power Supply Module
  • SIGMATEK 0332.554.03 Control Module 371071000154
  • SIGMATEK ETT312-E Touch Terminal 01-230-312
  • SIGMATEK CDM163 Control Module
  • SIGMATEK VSV046 16-023-046 Module
  • SIGMATEK AM221 20-017-221 Analog Module
  • SIGMATEK CTS051 12-053-051 Control Module
  • SIGMATEK AM221 20-017-221 Analog Module
  • Sigmatek CSDI161 12-891-161 Safety Input Module
  • Sigma Tek 4000C-1 Directional Gyro Indicator
  • Sigmatek ETV0501 12-230-0501 VARAN Terminal
  • Sigmatek DDM165 SLIDES Module
  • Sigmatek N100 20-011-100 Control Module
  • Sigmatek AKM54K-ANC2R-B0 PM Servo Motor
  • Sigmatek CTO166 Digital Output Module
  • Sigmatek CP212-K 20-004-212-K CPU Module
  • Sigmatek CDI161 Digital Input Module
  • Sigmatek DCC080 05-700-080-2 Control Module
  • SIGMATEK SCP011 S-DIAS Safety CPU
  • SIGMATEK SCP011 S-DIAS Safety CPU
  • SIGMATEK SRO022 Relay Output Module
  • SIGMATEK MDD 121 Drive Axis Module
  • SIGMATEK DSI021 SLIDES Siemens Interface
  • SIGMATEK DIAS DCP161 Module 05-004-161
  • Sigmatek ETT312 Built-in Touch Terminal
  • SIGMATEK DCP 643 DIAS Controller System
  • SIGMATEK CIO021 Multi I/O Module
  • SIGMATEK CTMS020 Control Module
  • Sigmatek CME 221 Memory Module
  • Sigmatek 0420.615.01 Control Board
  • Sigmatek CP111 20-004-111 CPU Module
  • Sigmatek 9405.065.03 Control Board
  • Sigmatek ETEK 01-450-032-K C-IPC 256MB Industrial PC
  • Sigmatek CTO163 12-007-163 Safety Output Module
  • Sigmatek TO081 20-007-081 Digital Output Module
  • Sigmatek SI021 20-022-021 Safety Input Module
  • Sigmatek 05-052-014 Control Module HW 1.0
  • Sigmatek TO127 20-007-127 Digital Output Module
  • Sigmatek SV141 20-023-141 VARAN Splitter
  • SIGMATEK CIV513 C-DIAS VARAN Control Module
  • Sigmatek DCP 642 SLIDES Control Module with DKL002
  • Sigma-Tek 5000B-36 Attitude Gyro
  • SIGMATEK 01-450-024 C-IPC Industrial Computer
  • SIGMATEK CP102 S-DIAS Processor Module
  • Sigmatek CRCH081 12-752-081 PLC Module
  • SIGMATEK SDD340-46 Series Drive
  • SIGMATEK TAE731 Touch Display Unit
  • SIGMATEK CDM167 Module 12-008-167-O
  • SIGMATEK TO161 Digital Output Module
  • Sigmatek NC4 45291000003 Control Board
  • Sigmatek HGW1033-3 Handheld Wireless Unit
  • Sigmatek CAI0811 Analog Input Module
  • Sigmatek STO081 S-DIAS Safety Output Module
  • Sigmatek C-IPC 01-450-171-V Industrial PC
  • Sigmatek S1 032-8AF61-R4 EZ Servo Motor
  • Sigmatek CM5V020 Baseplate 12-250-021-K
  • Sigmatek HGT833-W Panel
  • Sigmatek NC4 45291000003 Control Module
  • SIGMATEK 12-780-012 R8-IPC Geode LX800
  • SIGMATEK CCP081 Processor Module
  • SIGMATEK SCP011 S-DIAS Safety CPU
  • SIGMATEK ETT1533 Touchscreen 01-230-1533
  • SIGMATEK AI043 Analog Input Module
  • SIGMATEK CTMS020 Control Module
  • SIGMATEK SLIDES VARAN Interface VAC 012
  • SIGMATEK Compact Control Unit 06038666
  • SIGMATEK CP112 Control Module
  • SIGMATEK MDP 101-1 Power Supply Module
  • Sigmatek CCP621 C-SLIDES Processor Module
  • Sigmatek MDD111 Axle Module
  • Sigmatek CCL912 C-SLIDES CPU PSU Module
  • Demag NC4 0332.554.03 Sigmatek Control Board
  • Sigmatek Krauss Maffei PC321-K Control Cabinet PC
  • Sigmatek CNC W150 Circuit Board Wittmann Robot
  • Sigmatek AKM44G-ANCNGBB0 Servo Motor
  • Sigmatek CSCP011 Safety Module 12-890-011
  • Mannesmann Demag Sigmatek CP126 Central Unit
  • Sigmatek ISE021 Interface Module 20-101-021
  • Sigmatek CCP251 C-DIAS Processor Module
  • SIGMATEK ET 352 Terminal
  • SIGMATEK SV141 Module 20-023-141
  • SIGMATEK NC-337 3 Axis Module
  • Sigmatek MDP 101 Power Module
  • SIGMATEK HGT834-W Teach Pendant
  • Sigmatek C-DIAS Processor Module CCL912
  • SIGMATEK CP627 Module
  • SIGMATEK PC 323-K Controller
  • SIGMATEK MDD121 Controller Module
  • SIGMATEK ICA111 Module
  • SIGMATEK MDD 121-1 Drive Axis Module
  • SIGMATEK M5V121 12-300-121 Krauss Maffei M5 Display Unit
  • SIGMATEK CTMS 020 CD Technology Module 12-250-011
  • SIGMATEK PCT122 Touch Panel
  • Stöger MCS4100 SIGMATEK Control Module
  • SIGMATEK Krauss Maffei M5 Keyboard Unit M5K902 12-310-902
  • SIGMATEK SD310-2 Servo Drive Module
  • SIGMATEK CM5V020 CD Wiring Circuit Module
  • SIGMATEK ENGEL Robot Drive M3 02220-8748
  • Sigmatek CFF011 Drive Module
  • Sigmatek DCP642 DIAS Central Unit
  • Sigmatek SDM081 S-DIAS Safety Module
  • Sigmatek C-IPC 01-450-024 Industrial PC
  • Sigmatek Kompakt NC4 Control Module
  • Sigmatek VI021 VARAN Interface Module
  • Sigmatek Engel RD-M6-2 Robot Drive
  • Sigmatek CNC031 Positioning Module
  • Sigmatek TO161 Digital Output Module
  • Sigmatek CCL912 CPU PSU Module
  • Sigmatek ETV0821 Operator Terminal
  • SIGMATEK ET 261 Operator Terminal
  • SIGMATEK CP731-K Control Module 20-004-731-K
  • SIGMATEK ETEK Motherboard 12-250-021-K CM5V020
  • Sigmatek DEZ 181 DIAS Input Output Module
  • Sigmatek IPC111-D Industrial PC
  • SIGMATEK AKM54L-ANC2GBB0 Servomotor with Gearhead
  • SIGMATEK CTMS020 CM5V020 CNC Controller