Welcome to the Industrial Automation website!

NameDescriptionContent
XING-Automation
E-mail  
Password  
  
Forgot password?
  Register
当前位置:

Troubleshooting of R&S EPL1000

F: | Au:FANS | DA:2026-05-26 | 675 Br: | 🔊 点击朗读正文 ❚❚ ▶ | Share:


Time domain scanning configuration and performance failures

2.1 Spectrum leakage or low amplitude in time-domain scanning results

Fault phenomenon:

When using FFT based time-domain scanning to measure CISPR Band B (150kHz~30MHz), compared with traditional stepped scan results, the amplitude of certain frequency points is more than 2dB lower or non-existent false signals appear.

Principle Review:

The time-domain scanning of EPL1000 adopts parallel FFT processing to collect the entire CISPR frequency band at once, fully meeting the requirements of CISPR 16-1-1 for measurement time and detector response. However, FFT processing is sensitive to the signal-to-noise ratio and aliasing effects of the input signal. If the pre selection filter is not correctly enabled, it may lead to spectral leakage.

Solution:

Check the status of the pre selection filter: EPL1000 has built-in pre selection filters for CISPR Band A (9kHz~150kHz) and Band B (150kHz~30MHz). Under the 'FREQ' menu, confirm that 'Preselection' is set to 'Auto' or 'Band B'. If set to 'Off', strong signals can cause ADC saturation and harmonic distortion.

Adjust attenuator settings: Press the [AMPT] key to set the RF attenuation to "Auto". The pulse protection input of EPL1000 can withstand high instantaneous power, but excessive attenuation will reduce sensitivity. The automatic attenuation function optimizes the dynamic range based on the signal level.

Check overload indicator: If a red warning of "OVLD" appears in the status bar at the top of the screen, it indicates that the input signal exceeds the linear range of the mixer. At this point, it is necessary to increase the attenuation (manually set to 10dB or higher), or enable a 20dB preamplifier (only used in weak signals).

Measurement time setting: The measurement time for time-domain scanning should be at least 1 second to ensure that the CISPR quasi peak detector reaches steady state. If the setting is too short (such as 10ms), the quasi peak response will be insufficient and the amplitude will be lower. Recommended settings: Band B measurement time ≥ 1s, Band A measurement time ≥ 5s.

Verification method:

Comparative test: Under the same signal source (such as a comb wave generator), time-domain scanning and step scanning measurements should be used separately, and the deviation between the two should be within the ± 1.5dB range allowed by CISPR 16-1-1. If exceeded, contact the R&S service center for calibration.

2.2 Intermittent interference cannot be captured

Fault phenomenon:

There is occasional transient interference in the device (such as pulses generated by relay jumps), but multiple repeated scans have not shown this interference, resulting in missed reports in the pre compatibility test.

Reason:

The default Preview Scan uses a peak detector for fast scanning, but the scanning time is short (such as 100ms), which may miss transient events with low probability of occurrence.

Solution - Long term gapless measurement:

The time-domain scanning of EPL1000 supports a continuous measurement time of up to 100 seconds. The operation steps are as follows:

Enter the "Scan Table" menu and define the frequency range to be monitored (e.g. 150kHz to 30MHz).

Set the 'Measurement Time' to 10 seconds or higher (maximum 100 seconds). Note: The longer the measurement time, the larger the amount of data in a single scan, but it can ensure the capture of all transient signals.

Enable the 'Max Hold' tracking feature to overlay the maximum hold curve after multiple scans.

If you need to accurately locate the time point where interference occurs, use the Real time Spectrogram function - this mode ensures seamless acquisition of the entire frequency band and displays amplitude changes over time in color coded format.

case

When testing the laptop power adapter, it was found that some frequency points occasionally exceeded the CISPR 22 Class B limit. After setting the measurement time to 10 seconds and repeating the scan 5 times, the maximum retention curve clearly shows that the peak appears at 22.3 MHz. Further use of real-time spectrograms identified interference occurring every 2 seconds for approximately 50ms, ultimately confirming it as switch noise from the power management IC.

Troubleshooting of detector and limit line faults

3.1 Significant deviation between peak measurement results and certified laboratory

Fault phenomenon:

Under the same device and LISN, the quasi peak value (QP) measured by EPL1000 is more than 3dB lower than that of a third-party certified laboratory, but the peak detector results are consistent.

Root cause:

The charging and discharging time constants of the quasi peak detector strictly follow CISPR 16-1-1. If the measurement time (dwell time per point) of EPL1000 is insufficient, or the pulse repetition frequency of the measured signal is extremely low, the QP detector cannot reach a fully charged state.

Standard operation:

Confirm measurement time: For QP detectors, CISPR requires that the minimum measurement time for each frequency point depends on the intermediate frequency bandwidth of the receiver. For a bandwidth of 9kHz (Band B), the minimum dwell time is 1s; for a bandwidth of 200Hz (Band A), the minimum is 5s. In EPL1000, enter the "Detector" setting, set "Meas Time" to "Auto" or manually enter the above value.

  • LTI Motion CDE32.003.C3.0 Servo Positioning Controller
  • LTI DRIVES CDE34.010 D2.4 Servo Drive Unit
  • LUST LTi Drives CDD34.005.C2.1 Servo Drive
  • LTi Drives SO84.012.0002.0100.2 Servo Controller
  • LUST LTi Drives CDE34.010.D2.4 Servo Drive
  • LUST LTI Drives CDF30.008 Servo Drive
  • LTi Drives SO CM-P.0010.11.00.0 Power Module
  • LTi DRiVES SO84.008.0002.0100.2 Servo Controller
  • LTI MOTION Servo Drive SO22.006.0040.1000.1
  • LTi DRiVES SO24.007.0030.1000.0 ServoOne Junior
  • LTi DRiVES LUST BC1300 Braking Module
  • LUST LTI DRIVES CDE34.010 Variable Frequency Drive
  • LUST LTI DRIVES CDA32.004 Servo Drive
  • LTi DRiVES LST-074-2-30-560/T1 Servo Motor
  • LUST LTI DRIVES CDB32.008 Servo Drive Module
  • Lust Antriebstechnik LTI Drives VF1204S Inverter
  • Lust LTI Drives CDE32.008 C3.0 Positioning Controller
  • LTI Drives LUST Servo One C-5633 D-35633
  • Lust LTI Drives CDA34.008.C1.3 Servo Controller 3kW
  • LTI CDF30.008 C2.6 Servo Drive 10962161
  • LUST LTi Drives CDB32.004 C2.3 SH Servo Drive
  • Lust LTi Drives CDE34.010 D2.0 Servo Drive
  • LTi Motion CDA32.006 Servo Drive with CM-CAN1
  • Lust LTI Drives CDB32.008 C2.3 Servo Controller
  • LTi CDE34.017.W3.0 Servo Drive
  • LUST LTI LST-074-2-30-320/T1 Servomotor
  • Lust LTi CDF30.002 C3.0 Positioning Controller
  • LTi DRIVES CDF30.008.C2.4 Drive
  • LUST LTI CDB32.008 W2.4 BR PC1 Drive
  • LTi SO84.004.0002.0100.2 Servo Controller
  • LTi SO84.012.0030.0011.2 ServoOne Drive
  • LTI Drives LUST Servo One C-5633 D-35633
  • LTi DRiVES c-line Smart Card A028.V01.1
  • LUST LTI CD A34.008 W1.4 BR Drive
  • LUST LTI LSH-097-2-30-560/T1 Servomotor
  • LUST LTI LST-074-2-30-320/T1 Servomotor
  • LUST LTi CDB32.004 C2.4.SH Servo Drive
  • LTI Drives PM-E40.2DRA054P Pitch Inverter
  • Lust LTI Drives CDF30.008 Servo Drive 8673548
  • LTI Drives CDF30.002.C0.7 Servo Controller
  • Lust LTI Drives CDB32.008.W2.4 Servo Drive
  • Lust LTI Drives CDB34.003 Servo Motor Driver
  • Lust LTI Drives CDA32.004 Servo Drive 3084456
  • Lust LTI Drives CDE34.005.W2.2 Servo Drive
  • Lust LTI Drives CDA34.006.W3.0 Servo Drive
  • LTI Drives CDF30.008.C2.4 Servo Controller
  • LTI Drives Lust CDB32.004.C2.4.SH Servo Drive
  • Delta Design 1666339-501 Micro Stepper PCB 114970
  • APEXEL 100X Microscope Mobile LED Pocket Lens
  • Micro Innovation XV-152-D4-10TVR-10 Touch Screen
  • Micro Innovation XVH330-57MPI-1-10 Touch Screen Glass
  • Delta Design 1666339-501 Micro Stepper PCB 114967
  • Micro Design MC100-MA V3 Board
  • Delta Design 1666339-501 Micro Stepper PCB 114968
  • Delta Design 1666339 PCB Micro Stepper ITR 116487
  • Micro Innovation XV-440-10TVB-1-10 Panel
  • Sterling 9B21-3 Micro Switch Durable Design
  • ATTINY104-XNANO Microchip AVR Prototype Board
  • Delta Design 1666339 Micro Stepper PCB ITR 116539
  • Super Micro X7SBJ Industrial Main Board
  • Delta Design NMS BD 1698437-501 Stepper Board
  • Micro Innovations Turbo Track Keyboard KB420BL
  • Micro Design PC-1 PC-1-H-DU010 Operator Panel
  • TCI MICRO MD 250mA 7W Dimmable LED Driver
  • Delta Design 1666339 Micro Stepper PCB ITR 116485
  • Delta Design NMS BD 1698437-501 Stepper Board
  • Life Instruments 714-1120-1U Micro Curette
  • Delta Design 1657925-501 Micro Processor PCB
  • Micro 100 RTC-125-2 Carbide Engraving Tool
  • Micro Design Inc. 2106022 Liquid Level Sensor PCB
  • Periodontal Micro Surgical Full Kit 20pc Blue Titanium K1004
  • Micro Innovation XN-2DO-24VDC-0.5A Module
  • Wide Innovative Technology PS-01 Micro Positioner Set
  • HU FRIEDY NH5020 Castroviejo Needle Holder Set
  • Moeller Micro Innovation XV-232-57BAS-1-10 Touch Screen
  • Micro Design System MCVP-300VR Control Module 24VDC
  • Scanlan 3003-234-1 Titanium Jacobson Needle Holder
  • Micro Innovation XV-102-B0-35TQR-10-PLC Panel
  • Promot Micro Design 3x8 Output 2187F
  • Micro Innovation LIC-SW-Galileo 91 30 000600 License Certificate
  • Micro Design AI0486.A4-50 Control Module Board
  • Micro Innovation XN-2DO-24VDC-0.5A Output Module
  • Micro Design Not-Aus Relais AE 8.96 AE 896
  • Micro Innovation CPC-600M-10 Ferag Touch Panel HMI
  • Micro Design System 96-002-09/1 Adapter 18 Bit
  • Micro Design LK3160 AI0486 Motherboard
  • Micro Design System 96-002-01/2 GF-1 CPU Board
  • Micro Design Systems CF1-10TVA-100 Micro Panel
  • Eaton Micro Innovation XP-702-C0-15TXI-10 Panel
  • HU FRIEDY Castroviejo NH5020 Needle Holder
  • MPM Micro Precision 3.BMT100 P02 Balancing Station
  • Micro Innovation GF1-10TVD-101 Touch Panel
  • Micro Innovation XV-102-B0-35TQR-10-PLC
  • Micro Innovation XP-601-15T-10 Touch Panel
  • Micro Innovation License Certificate 9130000650 LIC-SW-MXPRO
  • Micro Innovation XVC-601-GTI-15-V1-000 Panel
  • MI Micro Innovation HPT-100-82FGBA-01 Panel
  • Micro Innovation XV-102-B3-35TQR-10 Panel Cable
  • Micro Innovation MC2-440-10TVB-1-2 HMI Panel
  • Promot Micro Design 2x8 Output 2087F
  • Micro Innovation XN-16DO-24VDC-0.5A Output Module
  • Micro Design System 96-002-03/1 Power Supply Board
  • Touch Screen for Moeller XV-252-57CNN-1-10 Digitizer
  • Micro Design Systems 96-002-03/1 Power Supply Board
  • Micropipette Variable Range 100-1000 Microliter
  • Murphy Nerve Hook 9 Inch Neurosurgery Instrument
  • Micro Innovation XN-2DI-24VDC-P Input Module 225169
  • Micro Innovation ST-GF0-57E0D-000 Touch Panel
  • Micro Innovation XN-S4T-SBBC Base Module
  • Micro Innovation XN-2DO-24VDC-0.5A-P Output
  • Micro Innovation XN-PF-24VDC-D 8550225186
  • Eaton Micro Innovation XV-230-57MPN-1-10 Panel
  • Micro Design MC100-BED REV2 ME96049 Board
  • Micro Innovation XN-PF-24VDC-D 85 50 225186
  • Micro Innovation XN-PF-24VDC-D Communication Module
  • Micro Innovation CPC-600M-10 FERAG Panel
  • Micro Design Systems 96-002-05/1 P-Bus Adapter
  • Micro Design System MPB1-TP 97-003-01/4 PCB Card
  • Eaton Micro Innovation XVS-440-12MPI-1-10 HMI
  • Micro Components 900806 Test Circuit Board
  • Promot Micro Design PL59-MOC ME95035
  • Micro Innovation XV-440-10TVB-1-21 Panel
  • Micro Forceps 3003-934 Micro Surgical Instrument
  • Eaton Micro Innovation XN-32DI-24VDC-P Module
  • MICRO INNOVATION AG XN-32DI-24VDC-P Input Module
  • Micro Innovation XN-PLC-CANOPEN 85 50 274124 Controller VN01-01
  • Micro Innovation XN-S4S-SBBS Base Module
  • Micro Innovation XN-2DO-24VDC-0.5A-P Module