Technical specifications: Detailed explanation of performance parameters in all dimensions
(1) Core specifications for conductivity measurement (AX41x/AX45x)
Range: Programmable from 0-0.5 to 0-10000 μ S/cm (adjusted according to electrode constant), with a minimum range of 10 times electrode constant and a maximum range of 10000 times electrode constant;
Measurement units: μ S/cm, μ S/m, mS/cm, mS/m, M Ω - cm, TDS;
Accuracy: better than ± 0.01% range within the range of 0-100 μ S/cm, and better than ± 1% reading within the range of 10000 μ S/cm;
Working temperature: -10 to 200 ℃ (14 to 392 ° F);
Temperature compensation: -10 to 200 ℃, programmable temperature coefficient 0-5%/℃, built-in NaCl/HCl/NH3/UPW/linear compensation curve;
Temperature sensor: Pt100/Pt1000 (programmable);
Reference temperature: 25 ℃ (77 ° F).
(2) AX411 Calculation Variable Specification
Ratio: 0-19999;
Difference: 0-10000 μ S/cm;
Passing rate/removal rate: 0-100.0%;
Total dissolved solids (TDS): 0-8000ppm;
Calculated pH: 7.0-10.0 pH for ammonia system and 7.0-11.0 pH for sodium hydroxide system (calculated according to Appendix VGB 450L 1988).
(3) AX416 pH/ORP measurement specifications
Input: pH/mV input+solution ground, Pt100/Pt1000/Alco 3k temperature sensor;
Input resistance: glass electrode>1 × 10 ¹³ Ω, reference electrode 1 × 10 ¹³ Ω;
Range: pH -2 to 16, ORP -1200 to+1200mV;
Minimum range: any 2pH span or 100mV;
Resolution: 0.01pH;
Accuracy: 0.01pH;
Temperature compensation: automatic/manual Nernst compensation (-10 to 200 ℃), process solution compensation (coefficient -0.05 to 0.02%/℃);
Calibration mode: 1/2 point automatic calibration (supporting ABB/DIN/Merck/NIST/US Tech buffer), 2 user-defined buffer tables, single point process calibration.
(4) Display and logging function
Display type: Dual 5-digit 7-segment backlit LCD, 16 character dot matrix single line display;
Energy saving function: The backlight can be set to remain on or automatically turn off after 60 seconds of inactivity;
Logbook: Electronic recording of process events and calibration data (requires the installation of an optional board);
Real time clock: Provides timestamps for the logbook and automatic/manual functions (requires the installation of an optional board).
(5) Communication and control functions
Digital communication: Profibus DP (requires the installation of an option board, some configurations of the panel installation version are not supported);
AX410 PID control: supports P/PI/PID mode, analog/pulse width/pulse frequency output, positive/negative action, proportional band 0.1-999.9%, integration time 1-7200 seconds, differentiation time 0.1-999.9 seconds.
(6) Mechanical and Environmental Specifications
Installation and Dimensions
Wall mounted/pipeline installation: IP65 protection, dimensions 192mm × 230mm × 94mm, weight 1kg;
Panel installation: IP65 (front only), size 96mm × 96mm × 162mm, weight 0.6kg;
Cable entry: Standard 5/7 M20 cable glands, North American version 7 1/2-inch Hubble gland knock off holes.
power supply specifications
Voltage: 100-240V AC (50/60Hz, minimum 90V, maximum 264V), 12-30V DC;
Power consumption: 10W;
Insulation: 2kV RMS from mains to ground.
environmental parameters
Working temperature: -20 to 55 ℃ (-4 to 131 ° F);
Storage temperature: -25 to 75 ℃ (-13 to 167 ° F);
Working humidity: ≤ 95% RH (no condensation);
EMC: Compliant with EN61326, EN50081-2, EN50082-2 standards.
Authentication and Security
Safety certification: UL, CE (covering EMC and low voltage directives, compliant with EN61010);
General safety: EN61010-1, input/output overvoltage class II, pollution level 2;
Language configuration: Supports English, French, German, Italian, and Spanish.
(7) AX416 sensor cleaning function
Cleaning mode: continuous/pulse (1-second on/off);
Frequency: 5 minutes to 24 hours (15 minute step within 1 hour, 1 hour step after 1 hour);
Duration: 15 seconds to 10 minutes (15 second step within 1 minute, 1 minute step after 1 minute);
Recovery time: 30 seconds to 5 minutes (30 second step).

Installation and wiring: convenient on-site implementation
The wiring design of the AX400 series fully considers the convenience of on-site operation. The terminal layout for wall/pipe installation and panel installation versions is different, but both use modular terminal block design, divided into three terminal blocks: A, B, and C. The C terminal block is an analog option board terminal, Profibus DP、 The extended current output and relay are both connected through this terminal block.
The sensor wiring block with terminal block B as the core can adapt to the wiring requirements of conductivity and pH/ORP sensors, supporting dual sensor inputs (sensor A/B). The wiring of the temperature compensator can be adapted to a 2-wire Pt100/Pt1000/Alco system through jumper wires, and the solution ground (ground rod) is the standard wiring for pH/ORP measurement, ensuring measurement stability. The equipment is also equipped with independent grounding bolts, which can achieve reliable grounding and reduce the impact of on-site electromagnetic interference on measurements.
All wiring terminals are designed for easy access, and wiring can be completed without disassembling the instrument body. The wall mounted/pipe mounted version maintains IP65 protection for the electronic part even when the terminal compartment is opened, and can be operated with water on site, greatly improving installation and maintenance efficiency.