Case 2: USB-3488A frequently disconnects and reconnects
Reason: USB port power management has enabled 'selective suspension'.
Solution: In Device Manager ->Universal Serial Bus Controller ->USB Root Hub Properties ->Power Management, uncheck 'Allow computer to shut down this device to save power'.
Case 3: No response after executing ibwrt, waiting indefinitely
Reason: The NDAC line of a certain instrument remains low, indicating incomplete data acceptance.
Solution: Use an oscilloscope to detect the NDAC line. If a device is abnormal, isolate it separately for testing; If all instruments are normal, check the GPIB transceiver at the controller end (replaceable card test).
Maintenance and environmental precautions
Temperature range: working from 0 to 55 ° C, storage from -20 to 80 ° C, humidity from 10 to 90% (no condensation).
Cleaning: Regularly blow the dust on the board with compressed air, especially on the golden finger area, and use an eraser to gently wipe the oxide layer.
Battery warning: The board does not contain lithium batteries (unlike some handheld devices), there is no risk of explosion, but the internal firmware of the USB version is non-volatile and does not require battery replacement.