2. Physical and environmental characteristics
Description of Characteristics and Specifications
Net weight (including accessories)<110 g (0.24 lb) Lightweight design for easy handheld measurement
Temperature range: -15 ℃~+65 ℃ (+5 ℉~+149 ℉); Non work: -62 ℃~+85 ℃ (-80 ℉~+185 ℉) Suitable for laboratory and industrial environments
Humidity 5 cycles (120 hours), 95%~97% relative humidity meets Tektronix standard 062-2847-00 Class 3
Altitude<2000 meters to avoid the impact of high altitude and low pressure on insulation performance
When the input impedance curve frequency is 1Hz~10 ⁸ Hz, the impedance and phase characteristics are stable. Figure 3 provides a typical impedance curve
3. Certification and Compliance
The probe complies with multiple international safety standards to ensure compliance for use in different regions
European standards: EN 61010-1/A2 (Safety of measuring equipment), EN 61010-2-031:1994 (Handheld probe components);
North American Standard: UL 3111-1 (First Edition) CSA C22.2 No. 1010.1-92、CSA C22.2 No. 1010.2.031-94;
Pollution level: Pollution Degree 2 (prohibited from use in environments with conductive pollutants);
Overvoltage category: Supports CAT I/CAT II/CAT III (classified according to voltage levels, such as CAT III applicable to distribution systems).
Accessories and Replacement Parts
1. Standard accessories (included with probe)
Accessory Name Tektronix Part Number Function Description
Scalable hook probe 013-0107-07 connects wires/component pins to achieve hands-free measurement
Compensation adjustment tool 003-1433-02 Adjust probe compensation adjuster
When using multiple probes with color marker tape 016-0633-00, the correspondence between the marker probe and the oscilloscope channel
SMT KlipChip and grounding lead 196-3305-00 connect the probe to a small/hard to reach grounding point
Low inductance grounding component 343-1003-01 (grounding ring)+195-4240-00 (short wire) reduces the inductance of the grounding lead and reduces high-frequency signal distortion
Insulated crocodile clip grounding lead 196-3113-02 probe grounding connection to grounding reference point
Probe protective shell 204-1049-00 protects the probe tip from mechanical damage
2. Optional accessories (extra optional)
Accessory Name, Part Number, Applicable Scenarios
BNC to probe tip adapter 013-0226-00 converts BNC interface signals into probe measurable tip signals
50 Ω BNC to probe adapter 013-0227-00 compatible with BNC signal source with 50 Ω impedance
Micro to small probe adapter 013-0202-03 enables compact probes to be compatible with small probe accessories
IC Test Probes (10 Pack) 015-0201-07 Accurate Measurement of IC Pins
Probe holder 352-0351-00 fixes the position of the probe and is suitable for long-term measurement
SMT KlipChip set (20 pieces) SMG50 batch SMT (surface mount technology) component grounding connection
3. Main replacement parts (for repair and replacement)
Component Name Part Number Remarks
Probe body (including cable) P6139A probe core component, replace as a whole in case of malfunction
Probe tip component (10X) 206-0441-00 contains 8.0pF capacitor and 9M Ω resistor. Replace when the tip is damaged
Grounding ring 343-1003-01 is the core component of the low inductance grounding component
Replace the BNC plug (male) 131-3219-00 when the BNC interface connecting the probe to the oscilloscope is damaged