Module replacement and upgrade
XN-9000 supports progressive upgrades. Adding the XN analysis module can improve throughput, and XN-10 and XN-20 can be combined. Add SP-10 for automatic slide staining, DI-60 for digital morphological pre classification, TS-10 for sorting and archiving, and RU-20 for expanding reagent capacity. Optional applications RET, PLT-F, BF, HPC can be activated later without the need to replace the analyzer.
When replacing modules, check the interface, track, software version, reagent channel, power supply, and communication. Perform quality control and calibration verification after upgrading. If replacing the analysis module, check the sample allocation rules, LIS mapping, and result review rules. If replacing SP-10 or DI-60, check the smear rules and classification rules. If replacing TS-10, check the sorting rules and archiving location.
Reliability guarantee of results
Standardization of screening, staining, pre classification, and retesting rules. Indoor quality control, inter room quality evaluation, calibration verification. Abnormal alarm review, result review rules. Regular maintenance, record and archive. Personnel training, standard operating procedures. Through systematic assurance, XN-9000 can operate stably for a long time, providing reliable results for clinical practice.