Sampling rate: 4 loops 6 times/second (update time 0.167 seconds), 8 loops 3 times/second (0.333 seconds), 16 loops 1.5 times/second (0.667 seconds); Measurement resolution 0.006% (>14 bits), supports automatic zero and full-scale calibration.
Sensor range and accuracy:
Thermocouples: Type B (66-1760 ℃, ± 4.0 ℃), Type E (-200-787 ℃, ± 1.0 ℃), etc., with accuracy measured in a 25 ℃ environment (Type B is only valid in the 800-3200 ° F range).
RTD: Only 4/8 circuits are supported, RTD1 (-100-275 ℃, ± 1.1 ℃) and RTD2 (-120-840 ℃, ± 1.6 ℃) require factory pre installed scaling resistors.
(2) Power Supply and Certification
Power requirements: 15-24VDC (± 3VDC), maximum current 1A, no-load current 300mA.
Compliance certification: UL ®、 C-UL ® Certification (compliant with UL) ® 61010-1 standard), CE mark (compliant with EU EMC directive), RoHS design.
Order Information
The configuration needs to be selected according to the following dimensions, and the final model needs to be confirmed with a Watlow sales representative:
Number of circuits: 4/8/16 circuits.
Firmware type: standard/temperature control/enhanced/customized.
Digital I/O termination method: screw terminal (TB18)/large-scale termination (SCSI).
Accessories: Do you need TB50 terminal board, SCSI cable (3 feet, including optional right angle connector), serial communication cable (10 feet, DB-9 female/bare wire).
Serial communication jumpers: EIA/TIA-232, EIA/TIA-485 (not terminated), EIA/TIA-485 (terminated).
Special Input: If RTD or linear current/voltage input is required, factory pre installed scaling resistors are required (RTD is not supported for 16 circuit models).