ENOTEC ENSITU 7000 Oxygen Analyzer

Product positioning and core advantages

Continuous monitoring of flue gas oxygen content is key to improving efficiency and reducing emissions in combustion control of small boilers, biomass combustion furnaces, and industrial furnaces. Traditional extraction analysis systems require complex sampling pipelines, coolers, and instrument air, which not only require a large amount of maintenance but are also prone to blockage or corrosion due to high dust and high temperature conditions. The ENSITU 7000 oxygen analyzer launched by ENOTEC is an in-situ oxygen transmitter probe designed specifically for this pain point.

ENSITU 7000 adopts the zirconia measurement principle and directly inserts it into the flue for in-situ measurement. Its biggest highlight is that it does not require calibration or instrument air, and can be installed, powered on, and measured right out of the box. The probe can withstand flue gas temperatures up to 600 ° C and extremely high dust loads, with a response time of only about 0.5 seconds, enabling rapid and continuous oxygen monitoring. Whether it is a gas, oil, or coal-fired boiler, ENSITU 7000 can provide stable combustion control signals.

This product offers three insertion depths: 313 mm, 536 mm, and 996 mm. The flange connection complies with the DN65 PN6 standard and can be easily integrated into existing boiler concepts. The electronic unit is detachable and supports wall or cabinet installation, making it easy for on-site wiring and later maintenance. Combined with the ENOTEC Remote free application, users can complete all configurations, real-time data viewing, and historical data archiving through their smartphones or tablets.


Measurement principle and structural analysis

The core of ENSITU 7000 is the zirconia oxygen sensor. Zirconia exhibits oxygen ion conductivity at high temperatures, and when there is an oxygen partial pressure difference on both sides of the sensor, a millivolt signal proportional to the logarithm of the oxygen concentration is generated. One side of the sensor is the reference air (usually ambient air, 20.95% O ₂), and the other side is the measured gas in the flue. By measuring the mV signal and combining it with the Nernst equation, the oxygen content in the flue gas can be calculated.

Unlike extractive systems, ENSITU 7000 uses in-situ diffusion. The smoke enters the sensor chamber through a large filter at the end of the probe, which is equipped with a V-shaped protective cover to effectively block dust particles and ensure rapid gas diffusion. This design avoids sampling tube blockage and condensation problems, making it particularly suitable for high dust and high humidity combustion environments.

The probe body is made of 1.4571 (316Ti) stainless steel, which has excellent corrosion resistance and high temperature strength. The electronic unit and probe are integrated through standard connectors, and if necessary, the electronic unit can be removed and installed in a location away from high-temperature radiation. The insertion depth of the probe can be selected according to the size of the flue, ensuring that the sensor is located in a representative airflow area.


Key technical specifications and selection references

The main technical parameters of ENSITU 7000 are as follows:

Measurement principle: Zirconia, in-situ type

Probe material: 1.4571 (316Ti)

Insertion depth: 313 mm, 536 mm, 996 mm

Flange connection: corresponding to DN65 PN6

Response time: approximately 0.5 seconds

Power supply voltage: 48 VDC, maximum 5 A

Accuracy: ± 0.2% O ₂

Environmental temperature: -40 ° C to+70 ° C

Process gas pressure: 950-1050 mbar

Protection level: IP 66/NEMA 4X

Process gas temperature: up to 600 ° C

Interface: 4-20 mA, ENOTEC Remote

When selecting, special attention should be paid to the insertion depth. 313 mm is suitable for small boilers or smoke ducts with limited space; 536 mm is suitable for medium-sized boilers; 996mm is used for large furnaces or in situations where deeper airflow is required. If the flue gas temperature approaches 600 ° C, it should be ensured that the probe material and installation method meet the requirements. The process gas pressure is close to atmospheric pressure. If the negative pressure in the flue is large, the sealing performance needs to be confirmed.

The output of the electronic unit is a 4-20 mA standard signal, which can be directly connected to DCS, PLC or combustion controller. The ENOTEC Remote interface provides wireless configuration and diagnostic functions, and parameter settings can be completed without opening the casing.


Installation guide: From flange to electrical connection

The installation of ENSITU 7000 is extremely simple and can usually be completed within a few minutes. The following are the standard installation steps:

Confirm installation location: Choose a location with stable airflow and good representativeness in the flue. Avoid approaching bends, baffles, or areas with direct flames. Ensure that the sensor end is located near the center of the flue after inserting the probe.

Welding or fastening flanges: Drill holes on the flue wall and weld DN65 PN6 matching flanges to ensure that the flange surface is perpendicular to the flue wall and sealed well. If using existing flanges, it is necessary to confirm that the specifications match.

  • National Instruments PCI-6250 Multifunction DAQ Device
  • National Instruments PCI-6224 Multifunction DAQ Card
  • National Instruments cRIO-9081 CompactRIO Controller
  • National Instruments USB-6353 X Series DAQ Device
  • National Instruments NI-9223 Simultaneous Analog Input Module
  • National Instruments PXIe-8133 Embedded Controller
  • National Instruments NI-9403 Digital I/O Module
  • National Instruments PCI-6111E Multifunction I/O Card
  • National Instruments FP-1001 FieldPoint Network Module
  • National Instruments PCI-6517 Industrial Digital I/O Card
  • National Instruments USB-5680 Power Meter
  • National Instruments PXI-6562 Digital Module
  • National Instruments PCI-6254 DAQ Card
  • National Instruments NI 9266 Output Module
  • National Instruments NI 9862 CAN Module
  • National Instruments NI 9217 Input Module
  • National Instruments PCI-4472 DSA Card
  • National Instruments NI 9232 Vibration Module
  • National Instruments NI-9269 Output Module
  • National Instruments GPIB-120B Interface
  • National Instruments NI 9209 Input Module
  • National Instruments NI 9210 Input Module
  • National Instruments NI 9381 Module
  • National Instruments PXI-2570 Relay Module
  • National Instruments 777318140 Cable
  • National Instruments 781049-01 Module
  • National Instruments 779140-01 Module
  • National Instruments PXI-6115 DAQ Card
  • National Instruments cDAQ-9185 Chassis
  • National Instruments NI-9375 I O Module
  • National Instruments USB-6366 X Series DAQ Device
  • National Instruments USB-6361 X Series DAQ Device
  • National Instruments PXI-4496 Sound and Vibration Module
  • National Instruments PCI-6289 Multifunction DAQ Card
  • National Instruments PXI-8119 Quad-Core Controller
  • National Instruments NI-9206 Analog Input Module
  • National Instruments 779516-01 Ordering Part Number
  • National Instruments cDAQ-9132 CompactDAQ Controller
  • National Instruments 782401-01 Ordering Part Number
  • National Instruments PXI-8105 Embedded Controller
  • National Instruments USB-6361 Multifunction I/O Device
  • National Instruments SCXI-1314 Terminal Block
  • National Instruments PCI-6515 Industrial Digital I/O Card
  • National Instruments 779038-01 Ordering Part Number
  • National Instruments cRIO-9012 CompactRIO Controller
  • National Instruments PXI-8364 MXI-Express Interface
  • National Instruments PXIe-8360 Interface Card
  • National Instruments 196859B-01 Ordering Part Number
  • National Instruments PXIe-1073 5-Slot Chassis
  • National Instruments NI 9209 Voltage Input Module
  • National Instruments USB-5133 Digitizer
  • National Instruments PXI-1044 Chassis
  • National Instruments NI 9266 785046-01 Module
  • National Instruments USB-6216 DAQ Device
  • National Instruments NI 9266 Output Module
  • National Instruments NI 9230 Vibration Module
  • National Instruments PCI-6230 DAQ Card
  • National Instruments PXI-6123 DAQ Card
  • National Instruments PCI-6255 DAQ Card
  • National Instruments PXI-8433 Serial Module
  • National Instruments PXIe-4138 Source Measure Unit
  • National Instruments 781044-01 PCIe-6321 DAQ Part Number
  • National Instruments PXIe-6356 X Series DAQ Module
  • National Instruments PCI-6513 Industrial Digital Output Card
  • National Instruments 183345A-04 Ordering Part Number
  • National Instruments PCIe-6351 X Series DAQ Card
  • National Instruments PXI-4070 FlexDMM Module
  • National Instruments PXI-6225 Multifunction DAQ Module
  • National Instruments 781496-01 Ordering Part Number
  • National Instruments 779592-01 Ordering Part Number
  • National Instruments SCXI-1121 Module
  • National Instruments NI 9148 Chassis
  • National Instruments PXI-6255 DAQ Module
  • National Instruments USB-4065 DMM
  • National Instruments PXI-6221 DAQ Module
  • National Instruments PXIe-1065 Chassis
  • National Instruments NI 9210 Input Module
  • National Instruments 778948-01 Output Module
  • National Instruments cRIO-9014 Controller
  • National Instruments cDAQ-9171 Chassis
  • National Instruments USB-6363 DAQ Device
  • National Instruments NI 9467 GPS Module
  • National Instruments USB-6351 DAQ Device
  • National Instruments NI 6581 Adapter Module
  • National Instruments 781045-01 Output Module
  • National Instruments PCI-6225 DAQ Card
  • National Instruments TRC-8543 CAN Cable
  • National Instruments PXIe-1073 Chassis
  • National Instruments USB-6003 DAQ Device
  • National Instruments NI 9476 Output Module
  • National Instruments myRIO-1900 Embedded Controller
  • National Instruments NI-9263 Analog Output Module
  • National Instruments 158561A-01L Ordering Part Number
  • National Instruments 785024-01 Ordering Part Number
  • National Instruments PCI-6510 Industrial Digital I/O Card
  • National Instruments USB-6259 Multifunction DAQ Device
  • National Instruments USB-8452 I2C SPI Interface Device
  • National Instruments PXI-8432 Serial Interface Module
  • National Instruments PXIe-5644R Vector Signal Transceiver
  • National Instruments PXI-2569 Relay Module
  • National Instruments 781174-01 Module
  • National Instruments 782000-01 Module
  • National Instruments PXI-5152 Digitizer
  • National Instruments PXI-6284 DAQ Card
  • National Instruments PCI-1428 Frame Grabber
  • National Instruments PXI-1050 Chassis
  • National Instruments NI USB-8486 Fieldbus
  • National Instruments NI 9862 CAN Module
  • National Instruments NI 9146 FPGA Chassis
  • National Instruments PXI-8108 Controller
  • National Instruments NI SWB-2816A Matrix Module
  • National Instruments NI 9269 Analog Output Module
  • National Instruments PCI-6031E Multifunction DAQ Card
  • National Instruments PXI-6713 Analog Output Module
  • National Instruments NI 9216 RTD Input Module
  • National Instruments NI-9237 Strain Bridge Input Module
  • National Instruments PXI-5124 High Resolution Digitizer
  • National Instruments USB-6343 X Series DAQ Device
  • National Instruments PXIe-6536 Digital I/O Module
  • National Instruments CAL-4330 Calibration Fixture
  • National Instruments PXI-4110 Programmable Power Supply
  • National Instruments 781503-01 Input Module
  • National Instruments FP-AI-V10 Module
  • National Instruments PXI-5114 Digitizer Card
  • National Instruments PXIe-8398 MXI Module
  • National Instruments cRIO-9056 Controller
  • National Instruments PCI-6509 DIO Device
  • National Instruments cRIO-9053 Controller