Maintenance and troubleshooting of ENOTEC SILOTEC 8000 silo analyzer

F: | Au:FANS | DA:2026-09-22 | 368 Br: | 🔊 点击朗读正文 ❚❚ ▶ | Share:

Maintenance and troubleshooting of ENOTEC SILOTEC 8000 silo analyzer

System positioning: O2/COE analysis for bin smoldering monitoring

SILOTEC 8000 is an O2/COE analysis system designed specifically for storage containers such as silos, coal bunkers, and biomass silos. It continuously extracts gas from the top space of the silo, while measuring oxygen content and CO equivalent, for early identification of smoldering fires. When smoldering occurs in the silo, the generation of hydrocarbons and CO consumes oxygen, resulting in an increase in COe concentration and a decrease in O2 concentration. SILOTEC 8000 can quickly capture this change and provide critical signals for safety interlocking, inerting protection, or alarm.

The system consists of electronic unit SME-53, measuring cell SIL-KES-0001, gas sampling probe SIL-GSF-8000, pneumatic unit, and GRP protective box. The electronic unit can be installed in a safe zone, and the measuring cell and sampling probe can be installed on the top of the silo, suitable for ATEX Zone 22 area. The sampling probe can extend into Zone 20. The maximum working pressure of the system is 0.5 bar, the maximum process gas temperature is 100 ° C, and the ambient temperature is -20 ° C to+55 ° C.


Measurement Principles and Safety Boundaries

SILOTEC 8000 adopts in-situ extraction measurement. The sampling probe is equipped with a sintered metal filter with a filtration accuracy of 3 μ m, which can safely isolate the explosive environment of dust in the silo from the heating sensor. The extraction system generates a continuous sample gas flow rate through an air injector, with a fast reaction speed. If equipped with a blowback module, compressed air can be used regularly to blow back the filter and reduce blockage.

Simultaneously install O2 sensors and COe sensors in the measurement pool. The O2 sensor is based on the zirconia principle, with instrument air (20.95% O2) as the reference side and sample gas as the measurement side. The difference in oxygen partial pressure on both sides generates an mV signal, which is converted to oxygen concentration using the Nernst equation. The COe sensor adopts the MXP principle and can detect CO, H2, and CxHy hydrocarbons. The catalytic electrode oxidizes incompletely oxidized gas molecules, generating a voltage signal related to the concentration of COe.

The safety boundary must be strictly adhered to: the system must not be used near flammable gases, otherwise there is a risk of explosion. The COe sensor requires a minimum of 0.3% O2 to measure correctly, and accuracy decreases below this value. The surface temperature of the measuring pool and the inlet and outlet of the sample gas may exceed 100 ° C. Insulated gloves must be worn during maintenance, and the power supply to the electronic unit must be cut off first. The measuring tank shall not be disassembled until it has cooled to below 35 ° C. It is absolutely forbidden to directly connect the probe to the 230V main power supply, otherwise it will immediately damage the heating element.


Installation and wiring points

The installation of electronic units should avoid vibrations greater than 2g and maintain a minimum distance from adjacent objects: 200mm at the top and bottom, and 50mm on the side. The ambient temperature must be between -20 ° C and+55 ° C, with a protection level of IP66. The installation position should be convenient for observing the display screen and operating the keyboard, and it is recommended to install it at a visual height.

The shielding layer of the probe signal cable FEP-0007/8 can only be connected to the PE terminal at the electronic unit end, and the probe end must not be grounded. The maximum length of the cable is 50m, the minimum bending radius of FEP-0007 is 252mm, and FEP-0008 is 329mm. The installation temperature is -5 ° C to+50 ° C, and the operating temperature is -40 ° C to+90 ° C. The signal cable should intersect perpendicularly with the power cable to avoid parallel wiring. The ferrite magnetic ring must be installed according to the specified wiring harness, otherwise the CE certification will be invalidated.

Pneumatic cable FEP-0002 is used for reference air and test gas, with a minimum bending radius of 138mm. When connecting, use a support sleeve, clamping ring, and nut, tighten by hand first, and then use a wrench to tighten 1.25 turns. The sampling probe flange needs to be air tight welded, and the filter locking torque is 250Nm. The electronic unit PE must be connected to the bin PE potential by the customer.

When wiring, please note that the O2 measuring cell, thermocouple, heater, probe solenoid valve, and COe sensor correspond to different terminals and colors. Thermocouple compensation cables must use ENOTEC original cables, otherwise the measurement will be inaccurate. COe sensor wire number 11-14, O2 sensor signal wire is white/brown and orange, heater is black/blue/green yellow.


Initial power on and software configuration

Before powering on, check whether the system number is consistent, whether the nameplate voltage matches, whether the wiring is correct, whether the gas circuit is airtight, whether the flange leaks, and whether the on-site conditions are consistent with the data sheet.

  • National Instruments PCI-6250 Multifunction DAQ Device
  • National Instruments PCI-6224 Multifunction DAQ Card
  • National Instruments cRIO-9081 CompactRIO Controller
  • National Instruments USB-6353 X Series DAQ Device
  • National Instruments NI-9223 Simultaneous Analog Input Module
  • National Instruments PXIe-8133 Embedded Controller
  • National Instruments NI-9403 Digital I/O Module
  • National Instruments PCI-6111E Multifunction I/O Card
  • National Instruments FP-1001 FieldPoint Network Module
  • National Instruments PCI-6517 Industrial Digital I/O Card
  • National Instruments USB-5680 Power Meter
  • National Instruments PXI-6562 Digital Module
  • National Instruments PCI-6254 DAQ Card
  • National Instruments NI 9266 Output Module
  • National Instruments NI 9862 CAN Module
  • National Instruments NI 9217 Input Module
  • National Instruments PCI-4472 DSA Card
  • National Instruments NI 9232 Vibration Module
  • National Instruments NI-9269 Output Module
  • National Instruments GPIB-120B Interface
  • National Instruments NI 9209 Input Module
  • National Instruments NI 9210 Input Module
  • National Instruments NI 9381 Module
  • National Instruments PXI-2570 Relay Module
  • National Instruments 777318140 Cable
  • National Instruments 781049-01 Module
  • National Instruments 779140-01 Module
  • National Instruments PXI-6115 DAQ Card
  • National Instruments cDAQ-9185 Chassis
  • National Instruments NI-9375 I O Module
  • National Instruments USB-6366 X Series DAQ Device
  • National Instruments USB-6361 X Series DAQ Device
  • National Instruments PXI-4496 Sound and Vibration Module
  • National Instruments PCI-6289 Multifunction DAQ Card
  • National Instruments PXI-8119 Quad-Core Controller
  • National Instruments NI-9206 Analog Input Module
  • National Instruments 779516-01 Ordering Part Number
  • National Instruments cDAQ-9132 CompactDAQ Controller
  • National Instruments 782401-01 Ordering Part Number
  • National Instruments PXI-8105 Embedded Controller
  • National Instruments USB-6361 Multifunction I/O Device
  • National Instruments SCXI-1314 Terminal Block
  • National Instruments PCI-6515 Industrial Digital I/O Card
  • National Instruments 779038-01 Ordering Part Number
  • National Instruments cRIO-9012 CompactRIO Controller
  • National Instruments PXI-8364 MXI-Express Interface
  • National Instruments PXIe-8360 Interface Card
  • National Instruments 196859B-01 Ordering Part Number
  • National Instruments PXIe-1073 5-Slot Chassis
  • National Instruments NI 9209 Voltage Input Module
  • National Instruments USB-5133 Digitizer
  • National Instruments PXI-1044 Chassis
  • National Instruments NI 9266 785046-01 Module
  • National Instruments USB-6216 DAQ Device
  • National Instruments NI 9266 Output Module
  • National Instruments NI 9230 Vibration Module
  • National Instruments PCI-6230 DAQ Card
  • National Instruments PXI-6123 DAQ Card
  • National Instruments PCI-6255 DAQ Card
  • National Instruments PXI-8433 Serial Module
  • National Instruments PXIe-4138 Source Measure Unit
  • National Instruments 781044-01 PCIe-6321 DAQ Part Number
  • National Instruments PXIe-6356 X Series DAQ Module
  • National Instruments PCI-6513 Industrial Digital Output Card
  • National Instruments 183345A-04 Ordering Part Number
  • National Instruments PCIe-6351 X Series DAQ Card
  • National Instruments PXI-4070 FlexDMM Module
  • National Instruments PXI-6225 Multifunction DAQ Module
  • National Instruments 781496-01 Ordering Part Number
  • National Instruments 779592-01 Ordering Part Number
  • National Instruments SCXI-1121 Module
  • National Instruments NI 9148 Chassis
  • National Instruments PXI-6255 DAQ Module
  • National Instruments USB-4065 DMM
  • National Instruments PXI-6221 DAQ Module
  • National Instruments PXIe-1065 Chassis
  • National Instruments NI 9210 Input Module
  • National Instruments 778948-01 Output Module
  • National Instruments cRIO-9014 Controller
  • National Instruments cDAQ-9171 Chassis
  • National Instruments USB-6363 DAQ Device
  • National Instruments NI 9467 GPS Module
  • National Instruments USB-6351 DAQ Device
  • National Instruments NI 6581 Adapter Module
  • National Instruments 781045-01 Output Module
  • National Instruments PCI-6225 DAQ Card
  • National Instruments TRC-8543 CAN Cable
  • National Instruments PXIe-1073 Chassis
  • National Instruments USB-6003 DAQ Device
  • National Instruments NI 9476 Output Module
  • National Instruments myRIO-1900 Embedded Controller
  • National Instruments NI-9263 Analog Output Module
  • National Instruments 158561A-01L Ordering Part Number
  • National Instruments 785024-01 Ordering Part Number
  • National Instruments PCI-6510 Industrial Digital I/O Card
  • National Instruments USB-6259 Multifunction DAQ Device
  • National Instruments USB-8452 I2C SPI Interface Device
  • National Instruments PXI-8432 Serial Interface Module
  • National Instruments PXIe-5644R Vector Signal Transceiver
  • National Instruments PXI-2569 Relay Module
  • National Instruments 781174-01 Module
  • National Instruments 782000-01 Module
  • National Instruments PXI-5152 Digitizer
  • National Instruments PXI-6284 DAQ Card
  • National Instruments PCI-1428 Frame Grabber
  • National Instruments PXI-1050 Chassis
  • National Instruments NI USB-8486 Fieldbus
  • National Instruments NI 9862 CAN Module
  • National Instruments NI 9146 FPGA Chassis
  • National Instruments PXI-8108 Controller
  • National Instruments NI SWB-2816A Matrix Module
  • National Instruments NI 9269 Analog Output Module
  • National Instruments PCI-6031E Multifunction DAQ Card
  • National Instruments PXI-6713 Analog Output Module
  • National Instruments NI 9216 RTD Input Module
  • National Instruments NI-9237 Strain Bridge Input Module
  • National Instruments PXI-5124 High Resolution Digitizer
  • National Instruments USB-6343 X Series DAQ Device
  • National Instruments PXIe-6536 Digital I/O Module
  • National Instruments CAL-4330 Calibration Fixture
  • National Instruments PXI-4110 Programmable Power Supply
  • National Instruments 781503-01 Input Module
  • National Instruments FP-AI-V10 Module
  • National Instruments PXI-5114 Digitizer Card
  • National Instruments PXIe-8398 MXI Module
  • National Instruments cRIO-9056 Controller
  • National Instruments PCI-6509 DIO Device
  • National Instruments cRIO-9053 Controller