Welcome to the Industrial Automation website!

NameDescriptionContent
XING-Automation
E-mail  
Password  
  
Forgot password?
  Register
当前位置:

Maintenance and troubleshooting of ENOTEC SILOTEC 8000 silo analyzer

F: | Au:FANS | DA:2026-09-22 | 369 Br: | 🔊 点击朗读正文 ❚❚ ▶ | Share:

After turning on the power, the system prompts to select language, set date and time, enter TAG number, ENOTEC Remote password (if enabled), and probe cable length. The cable length must be accurately inputted, as incorrect length can lead to measurement deviation and even damage to the COe sensor. Then enter the heating stage: the O2 sensor is heated first, and after reaching 800 ° C, the COe sensor begins to heat up. The display screen shows the current temperature, set temperature, heating power, and waiting information.

In measurement mode, the screen displays O2 and COe measurement values, active range, trend bar, and limit alarm indicators. Three LEDs respectively indicate alarm (orange), maintenance (orange), and error (red). The keyboard has four soft keys and numeric keys. The system code is set to 0000 at the factory, so there is no need to enter it to enter the menu. For safety reasons, modifications should be made and saved properly. If the code is lost, only trained service personnel can perform a system reset.

Calibration menu and automatic calibration

SILOTEC 8000 supports 1-point calibration and 2-point calibration, which can be performed separately for O2, COe, or both simultaneously. One point calibration uses test air (20.95% O2) to determine the offset. Two point calibration uses test air and test gas to determine offset and slope. Calibration can be manually initiated in CAL-MENU, or automatically calibrated with ACAL timing or remotely triggered.

ACAL settings are only visible when enabled. The calibration method can be either 1-point (test air only) or 2-point (test air+test gas). The startup methods include: time triggered, time+digital input, and digital input. The digital input requires a 12-24V DC control voltage to be applied to the "calibration release" terminal. If the control voltage still exists after calibration, the calibration will immediately restart.

Before calibration, it is necessary to confirm that the test air and test gas flow rates are correct. The test air flow rate should be 60-80 l/h, and the sample gas extraction flow rate should be 40-80 l/h. For systems with traffic monitoring, traffic can be viewed in System Checks. Testing gas recommendation: O2 sensor should use 2.1% O2 to balance N2; COe sensor should use the gas specified in the test certificate. Test gas flow rate of 40-80 l/h, maximum inlet pressure of 3 bar.

During the calibration process, the display screen shows trend charts, current measurement values, flow rates, and progress bars. If the sensor signal is unstable, it will report "sensor calibration failure". At this time, the 'test gas application time' can be extended, with a default of 10 minutes at the factory and an adjustable range of 5-30 minutes. The pre blowing time is used to flush the sensor before 2-point calibration and is turned off by default. The delay time is used to maintain the final measurement value after calibration and display the trend.


Sensor replacement and measurement pool maintenance

Before replacing the COe sensor, the COe measurement must be set to OFF through software. Otherwise, the new sensor may be immediately damaged. Step: Turn off the COe measurement, cut off the system power, disconnect the gas circuit and electrical connections of the measurement cell, remove the insulation layer of the measurement cell, loosen the COe sensor lock nut, remove the clamping ring, O-ring, and retaining ring in sequence, and take out the old sensor. Be aware that the sensor may experience high temperatures.

When preparing a new sensor, insert the clamping ring, O-ring, retaining ring, and locking nut, paying attention to the inclined surface facing the O-ring. Insert the new sensor and tighten the locking nut. The system needs to be tightened again after running for 2 hours. Reconnect terminals 11-14, power on, and enable COe measurement. The software will ask if the sensor needs to be replaced, and after confirmation, enter the test certificate parameters: COe RH0, Offset, Alpha, Beta, Gamma, Delta, SO2 correction value. The RH0 value must be correct, incorrect values will immediately damage the sensor. If there is an input error, immediately close and restart COe measurement.

When replacing the O2 sensor or probe tube, first remove the measuring cell, take out the inner core, remove the filter head and test gas distributor, loosen the flange screws of the measuring cell, and remove the old sensor. Clean the flange, install new sensors and gaskets, reinsert the inner core, and ensure even spring pressure. Perform 2-point calibration under process conditions 24 hours after installation.

When maintaining the measuring pool, pay attention to the hot surface. Disconnect the power supply, remove the insulation layer, loosen the fixing screws of the tank body, and take out the measuring tank. If the COe sensor is removed for a long time, the guide tube must be sealed with a locking rod and sealing kit to prevent false air from entering. The filter inspection is carried out according to the process cycle, checking the sealing and locking nut, with a torque of 250Nm. The blowback module can be automatically triggered when the extraction flow rate is lower than the alarm value, or it can be started during calibration or sensor inspection. If the flow rate remains low for about 5 seconds after blowback, the system enters an error state.

  • National Instruments PCI-6250 Multifunction DAQ Device
  • National Instruments PCI-6224 Multifunction DAQ Card
  • National Instruments cRIO-9081 CompactRIO Controller
  • National Instruments USB-6353 X Series DAQ Device
  • National Instruments NI-9223 Simultaneous Analog Input Module
  • National Instruments PXIe-8133 Embedded Controller
  • National Instruments NI-9403 Digital I/O Module
  • National Instruments PCI-6111E Multifunction I/O Card
  • National Instruments FP-1001 FieldPoint Network Module
  • National Instruments PCI-6517 Industrial Digital I/O Card
  • National Instruments USB-5680 Power Meter
  • National Instruments PXI-6562 Digital Module
  • National Instruments PCI-6254 DAQ Card
  • National Instruments NI 9266 Output Module
  • National Instruments NI 9862 CAN Module
  • National Instruments NI 9217 Input Module
  • National Instruments PCI-4472 DSA Card
  • National Instruments NI 9232 Vibration Module
  • National Instruments NI-9269 Output Module
  • National Instruments GPIB-120B Interface
  • National Instruments NI 9209 Input Module
  • National Instruments NI 9210 Input Module
  • National Instruments NI 9381 Module
  • National Instruments PXI-2570 Relay Module
  • National Instruments 777318140 Cable
  • National Instruments 781049-01 Module
  • National Instruments 779140-01 Module
  • National Instruments PXI-6115 DAQ Card
  • National Instruments cDAQ-9185 Chassis
  • National Instruments NI-9375 I O Module
  • National Instruments USB-6366 X Series DAQ Device
  • National Instruments USB-6361 X Series DAQ Device
  • National Instruments PXI-4496 Sound and Vibration Module
  • National Instruments PCI-6289 Multifunction DAQ Card
  • National Instruments PXI-8119 Quad-Core Controller
  • National Instruments NI-9206 Analog Input Module
  • National Instruments 779516-01 Ordering Part Number
  • National Instruments cDAQ-9132 CompactDAQ Controller
  • National Instruments 782401-01 Ordering Part Number
  • National Instruments PXI-8105 Embedded Controller
  • National Instruments USB-6361 Multifunction I/O Device
  • National Instruments SCXI-1314 Terminal Block
  • National Instruments PCI-6515 Industrial Digital I/O Card
  • National Instruments 779038-01 Ordering Part Number
  • National Instruments cRIO-9012 CompactRIO Controller
  • National Instruments PXI-8364 MXI-Express Interface
  • National Instruments PXIe-8360 Interface Card
  • National Instruments 196859B-01 Ordering Part Number
  • National Instruments PXIe-1073 5-Slot Chassis
  • National Instruments NI 9209 Voltage Input Module
  • National Instruments USB-5133 Digitizer
  • National Instruments PXI-1044 Chassis
  • National Instruments NI 9266 785046-01 Module
  • National Instruments USB-6216 DAQ Device
  • National Instruments NI 9266 Output Module
  • National Instruments NI 9230 Vibration Module
  • National Instruments PCI-6230 DAQ Card
  • National Instruments PXI-6123 DAQ Card
  • National Instruments PCI-6255 DAQ Card
  • National Instruments PXI-8433 Serial Module
  • National Instruments PXIe-4138 Source Measure Unit
  • National Instruments 781044-01 PCIe-6321 DAQ Part Number
  • National Instruments PXIe-6356 X Series DAQ Module
  • National Instruments PCI-6513 Industrial Digital Output Card
  • National Instruments 183345A-04 Ordering Part Number
  • National Instruments PCIe-6351 X Series DAQ Card
  • National Instruments PXI-4070 FlexDMM Module
  • National Instruments PXI-6225 Multifunction DAQ Module
  • National Instruments 781496-01 Ordering Part Number
  • National Instruments 779592-01 Ordering Part Number
  • National Instruments SCXI-1121 Module
  • National Instruments NI 9148 Chassis
  • National Instruments PXI-6255 DAQ Module
  • National Instruments USB-4065 DMM
  • National Instruments PXI-6221 DAQ Module
  • National Instruments PXIe-1065 Chassis
  • National Instruments NI 9210 Input Module
  • National Instruments 778948-01 Output Module
  • National Instruments cRIO-9014 Controller
  • National Instruments cDAQ-9171 Chassis
  • National Instruments USB-6363 DAQ Device
  • National Instruments NI 9467 GPS Module
  • National Instruments USB-6351 DAQ Device
  • National Instruments NI 6581 Adapter Module
  • National Instruments 781045-01 Output Module
  • National Instruments PCI-6225 DAQ Card
  • National Instruments TRC-8543 CAN Cable
  • National Instruments PXIe-1073 Chassis
  • National Instruments USB-6003 DAQ Device
  • National Instruments NI 9476 Output Module
  • National Instruments myRIO-1900 Embedded Controller
  • National Instruments NI-9263 Analog Output Module
  • National Instruments 158561A-01L Ordering Part Number
  • National Instruments 785024-01 Ordering Part Number
  • National Instruments PCI-6510 Industrial Digital I/O Card
  • National Instruments USB-6259 Multifunction DAQ Device
  • National Instruments USB-8452 I2C SPI Interface Device
  • National Instruments PXI-8432 Serial Interface Module
  • National Instruments PXIe-5644R Vector Signal Transceiver
  • National Instruments PXI-2569 Relay Module
  • National Instruments 781174-01 Module
  • National Instruments 782000-01 Module
  • National Instruments PXI-5152 Digitizer
  • National Instruments PXI-6284 DAQ Card
  • National Instruments PCI-1428 Frame Grabber
  • National Instruments PXI-1050 Chassis
  • National Instruments NI USB-8486 Fieldbus
  • National Instruments NI 9862 CAN Module
  • National Instruments NI 9146 FPGA Chassis
  • National Instruments PXI-8108 Controller
  • National Instruments NI SWB-2816A Matrix Module
  • National Instruments NI 9269 Analog Output Module
  • National Instruments PCI-6031E Multifunction DAQ Card
  • National Instruments PXI-6713 Analog Output Module
  • National Instruments NI 9216 RTD Input Module
  • National Instruments NI-9237 Strain Bridge Input Module
  • National Instruments PXI-5124 High Resolution Digitizer
  • National Instruments USB-6343 X Series DAQ Device
  • National Instruments PXIe-6536 Digital I/O Module
  • National Instruments CAL-4330 Calibration Fixture
  • National Instruments PXI-4110 Programmable Power Supply
  • National Instruments 781503-01 Input Module
  • National Instruments FP-AI-V10 Module
  • National Instruments PXI-5114 Digitizer Card
  • National Instruments PXIe-8398 MXI Module
  • National Instruments cRIO-9056 Controller
  • National Instruments PCI-6509 DIO Device
  • National Instruments cRIO-9053 Controller