Periodic blasting tank blowing to remove dust accumulation inside the probe filter tube;
Water cooling options include water/water or gas/water;
The official commitment is that the kiln gas data has a 95% availability rate and only requires one maintenance cycle throughout the year.
Under the dual pressure of drastic fluctuations in RDF (waste derived fuel) calorific value and tightening of cement emission regulations in various countries year by year, this "never stick, power-off self-protection" probe design has become the industry standard configuration for rotary kiln combustion control. In addition, the ENOTEC system can be integrated to complete planning, installation, and debugging as needed, without being bound to specific brands, and can flexibly respond to personalized needs.
Selection Decision Tree: Clarify Four Questions
Based on the above parameters, the selection process can be summarized into four steps:
Measurement parameters: as long as O ₂ → OXITEC or ENSITU; Need for unburned molecule joint testing → COMTEC; Drying process → AQUATEC; Storage bin safety → SILOTEC.
Installation method: CEM cabinet centralized layout → OXITEC 500E extraction type; On site in-situ → probe type.
Explosion proof requirement: safe area → safe area; Gas explosion zone 1/2 → GasEx; Dust area 20/21/22 → DustEx or SILOTEC 8000.
Operation and maintenance conditions: no instrument wind, hope for mobile phone operation and maintenance → ENSITU 7000; Extreme kiln conditions → CEMTEC; High temperature>600 ℃ → must be equipped with a cooling protection tube version.