How to purchase Honeywell Sensepoint XCD?
Product overview
The Honeywell Sensepoint XCD is a fixed gas detector suitable for industrial applications, capable of detecting flammable gases, toxic gases, and oxygen. It has a 3-wire 4-20mA output and RS485 MODBUS communication function. It is equipped with built-in alarm and fault relays, a local display screen, and can be fully configured through a non-invasive magnetic switch interface to protect personnel and factories from related gas hazards.
Electrical specifications
Input voltage range: 12-32VDC (nominal 24VDC)
Maximum power consumption: Depending on the type of gas sensor used, the electrochemical cell is 3.7W, the infrared sensor is 3.7W, and the catalytic sensor is 4.9W; the maximum surge current at 24VDC is 800mA
Current output: can choose sinking or sourcing mode
Relays: 3, 5A@250VAC , can choose between normally open or normally closed (switch) and power on/off (programmable); Alarm relay defaults to normally on/off, fault relay defaults to normally on/on
Communication: RS485, MODBUS RTU
Structure specifications
Material: The shell is made of epoxy coated aluminum alloy ADC12 or 316 stainless steel; The sensor is made of 316 stainless steel
Weight (approximately): Aluminum alloy LM25 is 4.4lbs; 316 stainless steel weighs 11lbs
Installation: Integrated installation board with 4 mounting holes suitable for M8 bolts; Optional pipeline installation kit, suitable for horizontal or vertical pipelines (diameter 1.5-3 inches, nominal 2 inches)
Cable entry: UL/cUL version includes 2 3/4 inch NPT conduit entries, providing suitable plug closure (if only 1 entry is used), sealed to maintain IP rating; ATEX/IECEx version includes 2 M20 cable entry points
Environmental Specifications
IP rating: IP66 (compliant with EN60529:1992)
Certification temperature range: -40 º F to+149 º F (-40 º C to+65 º C)
Detectable gas and sensor performance
Gas type user selectable full-scale range default range user selectable step size default calibration point response time (T90) (seconds) accuracy working temperature range default alarm point A1 default alarm point A2
Authentication
United States, Latin America, Canada: UL/c-UL - Class I Zone 1 Groups B, C, D, Class I Zone 2 Groups B, C, D, Class II Zone 1 Groups E, F, G, Class II Zone 2 Groups F, G, -40 ° C to+65 ° C
European international: IEC Ex d IIC Gb T6 (ambient temperature -40 º C to+65 º C), Ex tb IIIC T85 ° C Db IP66; ATEX Ex II 2 GD Ex d IIC Gb T6 (ambient temperature -40 º C to+65 º C), Ex tb IIIC T85 ° C Db IP66
Electromagnetic compatibility: CE: EN50270:2006, EN6100-6-4:2007
Performance: UL508; CSA 22.2 No.152 (flammable gases, excluding infrared sensors); ATEX,IEC/EN60079-29-1:2007,EN45544,EN50104,EN50271; China: PA type approval (transmitters and toxic gas sensors), "CCCF" Shenyang (flammable gas, approved by the fire department)