(5) Troubleshooting

The LCD displays various fault messages, such as "low transmission", "laser alignment error", "PLC reading error", etc. Based on the message explanation, corresponding measures are taken, such as cleaning the window, checking the connection, adjusting the alignment, etc. If the problem cannot be solved, download the error log and system log and contact the service personnel.


Electrical connection

(1) Transmitter unit interface

There are receiver connectors, main power connectors, service connectors, and network connectors, which are used to connect receiver cables, power and sensor inputs, service PCs, and Ethernet, respectively.

(2) Receiver cable connection

Detailed wiring of the receiver cable at both ends, including power supply, signal, temperature sensor, etc.

(3) Power cable connection

The wiring at both ends of the power cable includes temperature, pressure, flow probe input, and power input.

(4) RS-232 and Ethernet connectors

The functions of each terminal of the RS-232 connector and the wiring of the Ethernet RJ-45 connector.

(5) Ethernet connection to AO2000

According to the AO2000 software version, the method of connecting one or more laser analyzers is introduced, including IP address setting and network topology structure.

(6) Current circuit (4-20mA) input connection

The connection method of active and passive 4-20mA current loop input probes was demonstrated, and multiple instruments can be connected in series to the same probe.

(7) Power cable connection

The power cable is connected to the power supply unit, and the cable diameter should be within the range of 5-10mm. The power terminal can accommodate conductors with an area of less than 2.5mm ².

(8) Transmitter board - fuse and LED

The transmitter board has a main fuse and some LEDs to indicate the status of different power supply voltages. If the LEDs are abnormal, the fuse and power supply need to be checked.

欢迎您光临兰州实华分析技术有限公司

Span and zero check options

(1) Span inspection with circulation pool

Used to verify the performance of the analyzer, non corrosive gases use a receiver unit with an internal volume of 100mm, corrosive gases use a dedicated span pool, initiate span mode through digital communication, flush the internal pool, measure span concentration, then flush residual gases and return to normal mode.

(2) Span and zero inspection with internal sealed pool

The receiver unit contains a sealed gas pool that is activated through digital communication for span and zero modes. The internal pool is moved to the laser beam path by a servo motor, and the absorption signal in normal mode is calculated and stored for subtraction during span inspection and measurement of span gas concentration.


Class I Division 2 Version Description

(1) Preliminary explanation

This version contains installation, operation, and maintenance information, requiring the use of Division 2 wiring methods, conduit sealing to prevent explosion spread, and following the installation instructions of the sealing compound manufacturer.

(2) Instrument Design

Composed of transmitter unit, transmitter connection box, receiver unit, and power supply box, the power supply box is not suitable for installation in Class I, Div. 2 areas.

(3) Product Data

Environmental conditions: working temperature -20 ° C to+55 ° C, same storage temperature, protection level IP65 (non certified part), pollution level 2, installation category II, altitude 2000 meters, humidity 0 to 95%.

Power supply: Input 100-240V AC, output 24V DC, transmitter unit input 18-36V DC, maximum 20W.

Interface: RS232 complies with relevant standards, Ethernet protocol 10/100Base-T.

CSA certification: applicable to Class I Division 2 Groups A, B, C, D, temperature code T4.

(4) Warning

failure to follow safety instructions may cause ignition of hazardous atmosphere, Causing serious personal injury and/or property damage, replacing parts may impair the applicability of Class I Division 2.

(5) Electrical installation

Safety precautions: The use of cable glands must comply with NEC standards and local regulations. Unused cable glands should be closed according to the manufacturer's instructions. Power supply units are not suitable for installation in Class I, Div. 2 areas. Service PCs should be placed in explosion-proof safety areas.

Electronic interface of transmitter unit: The receiver and main power supply Myers Hubs are defined by the factory, and communication can use one of the two Myers Hubs in the center.

Cable connection: including detailed wiring for receiver cables, power cables, RS232, and Ethernet connections.


ATEX Zone 2 Version Description

(1) Preliminary explanation

Contains installation, operation, and maintenance information for this version.

(2) Instrument Design

Composed of transmitter unit, receiver unit, and transmitter connection box, the power supply box is not suitable for installation in Zone 2 area.

(3) Product Data

Environmental conditions: working temperature -20 ° C to+55 ° C, same storage temperature, protection level IP66, pollution level 2, installation category II, altitude 2000 meters, humidity 0 to 95%.

Power supply: Input 100-240V AC, output 24V DC, transmitter unit input 18-36V DC, maximum 20W.

  • National Instruments PCI-6250 Multifunction DAQ Device
  • National Instruments PCI-6224 Multifunction DAQ Card
  • National Instruments cRIO-9081 CompactRIO Controller
  • National Instruments USB-6353 X Series DAQ Device
  • National Instruments NI-9223 Simultaneous Analog Input Module
  • National Instruments PXIe-8133 Embedded Controller
  • National Instruments NI-9403 Digital I/O Module
  • National Instruments PCI-6111E Multifunction I/O Card
  • National Instruments FP-1001 FieldPoint Network Module
  • National Instruments PCI-6517 Industrial Digital I/O Card
  • National Instruments USB-5680 Power Meter
  • National Instruments PXI-6562 Digital Module
  • National Instruments PCI-6254 DAQ Card
  • National Instruments NI 9266 Output Module
  • National Instruments NI 9862 CAN Module
  • National Instruments NI 9217 Input Module
  • National Instruments PCI-4472 DSA Card
  • National Instruments NI 9232 Vibration Module
  • National Instruments NI-9269 Output Module
  • National Instruments GPIB-120B Interface
  • National Instruments NI 9209 Input Module
  • National Instruments NI 9210 Input Module
  • National Instruments NI 9381 Module
  • National Instruments PXI-2570 Relay Module
  • National Instruments 777318140 Cable
  • National Instruments 781049-01 Module
  • National Instruments 779140-01 Module
  • National Instruments PXI-6115 DAQ Card
  • National Instruments cDAQ-9185 Chassis
  • National Instruments NI-9375 I O Module
  • National Instruments USB-6366 X Series DAQ Device
  • National Instruments USB-6361 X Series DAQ Device
  • National Instruments PXI-4496 Sound and Vibration Module
  • National Instruments PCI-6289 Multifunction DAQ Card
  • National Instruments PXI-8119 Quad-Core Controller
  • National Instruments NI-9206 Analog Input Module
  • National Instruments 779516-01 Ordering Part Number
  • National Instruments cDAQ-9132 CompactDAQ Controller
  • National Instruments 782401-01 Ordering Part Number
  • National Instruments PXI-8105 Embedded Controller
  • National Instruments USB-6361 Multifunction I/O Device
  • National Instruments SCXI-1314 Terminal Block
  • National Instruments PCI-6515 Industrial Digital I/O Card
  • National Instruments 779038-01 Ordering Part Number
  • National Instruments cRIO-9012 CompactRIO Controller
  • National Instruments PXI-8364 MXI-Express Interface
  • National Instruments PXIe-8360 Interface Card
  • National Instruments 196859B-01 Ordering Part Number
  • National Instruments PXIe-1073 5-Slot Chassis
  • National Instruments NI 9209 Voltage Input Module
  • National Instruments USB-5133 Digitizer
  • National Instruments PXI-1044 Chassis
  • National Instruments NI 9266 785046-01 Module
  • National Instruments USB-6216 DAQ Device
  • National Instruments NI 9266 Output Module
  • National Instruments NI 9230 Vibration Module
  • National Instruments PCI-6230 DAQ Card
  • National Instruments PXI-6123 DAQ Card
  • National Instruments PCI-6255 DAQ Card
  • National Instruments PXI-8433 Serial Module
  • National Instruments PXIe-4138 Source Measure Unit
  • National Instruments 781044-01 PCIe-6321 DAQ Part Number
  • National Instruments PXIe-6356 X Series DAQ Module
  • National Instruments PCI-6513 Industrial Digital Output Card
  • National Instruments 183345A-04 Ordering Part Number
  • National Instruments PCIe-6351 X Series DAQ Card
  • National Instruments PXI-4070 FlexDMM Module
  • National Instruments PXI-6225 Multifunction DAQ Module
  • National Instruments 781496-01 Ordering Part Number
  • National Instruments 779592-01 Ordering Part Number
  • National Instruments SCXI-1121 Module
  • National Instruments NI 9148 Chassis
  • National Instruments PXI-6255 DAQ Module
  • National Instruments USB-4065 DMM
  • National Instruments PXI-6221 DAQ Module
  • National Instruments PXIe-1065 Chassis
  • National Instruments NI 9210 Input Module
  • National Instruments 778948-01 Output Module
  • National Instruments cRIO-9014 Controller
  • National Instruments cDAQ-9171 Chassis
  • National Instruments USB-6363 DAQ Device
  • National Instruments NI 9467 GPS Module
  • National Instruments USB-6351 DAQ Device
  • National Instruments NI 6581 Adapter Module
  • National Instruments 781045-01 Output Module
  • National Instruments PCI-6225 DAQ Card
  • National Instruments TRC-8543 CAN Cable
  • National Instruments PXIe-1073 Chassis
  • National Instruments USB-6003 DAQ Device
  • National Instruments NI 9476 Output Module
  • National Instruments myRIO-1900 Embedded Controller
  • National Instruments NI-9263 Analog Output Module
  • National Instruments 158561A-01L Ordering Part Number
  • National Instruments 785024-01 Ordering Part Number
  • National Instruments PCI-6510 Industrial Digital I/O Card
  • National Instruments USB-6259 Multifunction DAQ Device
  • National Instruments USB-8452 I2C SPI Interface Device
  • National Instruments PXI-8432 Serial Interface Module
  • National Instruments PXIe-5644R Vector Signal Transceiver
  • National Instruments PXI-2569 Relay Module
  • National Instruments 781174-01 Module
  • National Instruments 782000-01 Module
  • National Instruments PXI-5152 Digitizer
  • National Instruments PXI-6284 DAQ Card
  • National Instruments PCI-1428 Frame Grabber
  • National Instruments PXI-1050 Chassis
  • National Instruments NI USB-8486 Fieldbus
  • National Instruments NI 9862 CAN Module
  • National Instruments NI 9146 FPGA Chassis
  • National Instruments PXI-8108 Controller
  • National Instruments NI SWB-2816A Matrix Module
  • National Instruments NI 9269 Analog Output Module
  • National Instruments PCI-6031E Multifunction DAQ Card
  • National Instruments PXI-6713 Analog Output Module
  • National Instruments NI 9216 RTD Input Module
  • National Instruments NI-9237 Strain Bridge Input Module
  • National Instruments PXI-5124 High Resolution Digitizer
  • National Instruments USB-6343 X Series DAQ Device
  • National Instruments PXIe-6536 Digital I/O Module
  • National Instruments CAL-4330 Calibration Fixture
  • National Instruments PXI-4110 Programmable Power Supply
  • National Instruments 781503-01 Input Module
  • National Instruments FP-AI-V10 Module
  • National Instruments PXI-5114 Digitizer Card
  • National Instruments PXIe-8398 MXI Module
  • National Instruments cRIO-9056 Controller
  • National Instruments PCI-6509 DIO Device
  • National Instruments cRIO-9053 Controller