Welcome to the Industrial Automation website!

NameDescriptionContent
HONG  KANG
E-mail  
Password  
  
Forgot password?
  Register
当前位置:

YOKOGAWA MY600 Digital Insulation Resistance Tester

来源: | 作者:FAN | 发布时间 :2025-11-03 | 464 次浏览: | 🔊 Click to read aloud ❚❚ | Share:



YOKOGAWA MY600 Digital Insulation Resistance Tester

Product positioning and core functions

MY600 is a compact digital insulation resistance tester designed to measure the insulation resistance of electrical equipment or circuits. It supports voltage measurement (AC/DC) and low resistance on/off inspection, and is portable, high-precision, and adaptable to multiple scenarios. It is widely used in fields such as power equipment maintenance, electrical installation acceptance, and industrial circuit testing.

Packaging and Accessories

1. Standard accessories (included with the product)

After opening the box, it is necessary to confirm that the following accessories are intact. If they are missing or damaged, please contact the Yokogawa distributor:

Accessory Name Model/Part Number Quantity Usage

Portable case 93045 1 for storing instruments and accessories, easy to carry

Line probe with remote switch 98008 1 connected to LINE terminal for line side measurement

Grounding probe group 98009 1 is connected to the EARTH terminal for measuring the grounding side

Shoulder strap 99018 1 hanging instrument, supports dual hand operation

Alkaline dry battery -4 instrument power supply (AA alkaline battery recommended)

User manual set IM MY600-01EN/92Z1/00C01C01-01Z2/PIM113-01Z2, each containing multilingual instructions and global contact information

2. Optional accessories (to be purchased separately)

According to the requirements of the measurement scenario, it is necessary to ensure that the accessories meet the rated parameters of the instrument:

Accessory Name Model/Part Number Usage

USB communication adapter 91030 is used to connect to a PC and transfer measurement data from memory

Hook probe tip 99012 replaces standard probe, suitable for special wiring scenarios

Long probe tip 99013 deep into narrow spaces for measurement, expanding measurement range

image.png

Safety regulations

1. Core safety warning (to avoid electric shock/equipment damage)

Laser and high voltage protection: When measuring insulation resistance, there is high voltage at the tip of the probe, and touching the probe or the tested circuit is prohibited; After measurement, wait for automatic discharge to complete(“ ⚠️”  Symbol flashing+buzzing), then touch the circuit.

Measurement category restriction: The rated measurement category of the instrument is CAT III 600V, and it is prohibited to use it for CAT IV or main power circuits exceeding 600V; When the probe is paired with different accessories, the category is different (for example, 98008 with insulation cap is CAT III 600V, and without cap is CAT II 1000V).

Requirements for the tested equipment: Before measurement, the power supply of the DUT must be cut off, and the voltage measurement function must be used to confirm that there is no power before starting; Do not measure live circuits to avoid instrument damage or electric shock.

Equipment status check: If the instrument casing is damaged, the probe cable is exposed, or the battery compartment is damp, use is prohibited; After falling or colliding, it is necessary to contact the dealer for maintenance to avoid insulation protection failure.

2. Operation safety rules

Battery replacement: Before replacing, power off and remove the probe. It is forbidden to open the battery compartment during measurement; New batteries of the same type should be used to avoid mixing old and new/different models.

Environmental restrictions: Do not use in flammable gas, humid (with condensation) or outdoor rainy environments; Working temperature range -10~+50 ℃, storage temperature -20~+60 ℃, avoid direct sunlight.

Probe usage: Only Yokogawa designated probes can be used, damaged or aged probes should be replaced in a timely manner; Before connecting/disassembling the probe, it is necessary to disconnect it from the device being tested.


Core functions and operating procedures

1. Measurement mode and parameters

The instrument supports three core measurement modes, and the parameters and applicable scenarios for each mode are as follows:

(1) Voltage measurement (AC/DC automatic detection)

Range and Range: Automatic range (300.0V/600V), measurement range AC 2.0~600V (45~65Hz), DC ± 2.0~± 600V, over range display ">629V" (AC/positive DC) or "<-629V" (negative DC).

Accuracy: ± 1% reading ± 4 digits (23 ℃± 5 ℃, RH ≤ 80%), AC detected using true RMS, non sinusoidal (CF<2.5) requires an additional ± 1% error.

Operation steps:

Connect the line probe (98008) to the LINE terminal, and connect the ground probe (98009) to the EARTH terminal.

Set the range switch to "V/Ω" and connect the probes to the line side and ground side of the tested circuit.

No need to press the measurement switch, the instrument automatically detects AC/DC and displays the value (triggering a live warning when ≥ 30V:“ ⚠️”  Blinking and buzzing.

(2) Insulation resistance measurement (core function)

Rated voltage and range: Supports six rated voltages of 50V/100V/125V/250V/500V/1000V, with automatic range switching (such as 1000V range of 4~4000M Ω), and a fixed value displayed for exceeding the range (such as>4199M Ω).

Key parameters:

Open circuit voltage: 100-110% of rated voltage.

Short circuit current: ≤ 1.5mA (1000V range) to ensure measurement safety.

Accuracy: First effective range (such as 1000V range 0.100~1000M Ω) ± 2% reading ± 2 digits; Second effective range (1001~4000M Ω) ± 5% reading.

Featured Features:

Automatic discharge: After measurement, keep the probe connected, and the instrument will automatically release the charge of the measured capacitor load. Before the discharge is completed“ ⚠️”  Flashing and buzzing.

Pass/Tail judgment: A reference value (0.001~4199M Ω) can be set. When the measured value is ≥ the reference value, the backlight flashes green and "PASS", otherwise it flashes red and "FAIL".

DAR/PI calculation: Automatically calculate the dielectric loss absorption ratio (DAR=1 minute value/15 second value) and polarization index (PI=10 minute value/1 minute value), ranging from 0.00 to 9.99, with ">9.99" displayed if out of range.

Operation steps:

Connect the probe, confirm that the tested device is powered off, and verify that there is no power using voltage measurement.

Turn the range switch to the corresponding rated voltage level (long press SELECT to switch between 125V/100V levels).

Connect the probe to the device under test, and simultaneously press the instrument's "measurement switch" and the probe's "remote switch" to start measuring (continuous measurement requires locking the measurement switch).

After the measurement is completed, wait for the automatic discharge to end before removing the probe.

(3) Low resistance measurement (on-off check)

Range and Range: Automatic Range (40.00 Ω/400.0 Ω/4000 Ω), Measurement Range 0.00-4199 Ω, Over Range Display ">4199 Ω".

Key parameters:

Open circuit voltage: DC 4~6.9V.

Measurement current: below 2 Ω ≥ 200mA, when the current is ≥ 200mA, a buzzer will sound to indicate on/off.

Accuracy: 0.20-4000 Ω± 2.5% reading ± 8 digits; 0~0.19 Ω± 8 digits (0 Ω calibration needs to be performed first).

Featured feature: 0 Ω calibration (can counteract probe and fuse resistance, up to 3 Ω), ensuring low resistance measurement accuracy.

Operation steps:

Turn the range switch to "V/Ω" and press SELECT to switch to low resistance mode.

Short circuit the probe, long press the "0 Ω ADJ" button to perform calibration (display "0.00 Ω" and light up the "0 Ω" indicator light).

Connect the probe to the tested circuit, press the measurement switch, and display the resistance value (beep when the current is ≥ 200mA).

2. Auxiliary functions

Backlight and LED lights: The illuminance sensor automatically detects the ambient brightness, automatically turns on the backlight and measurement point LED lights when dim, and automatically turns off after 2 minutes of inactivity; It can also be manually set to be normally off.

Automatic shutdown: Automatic shutdown after 10 minutes of inactivity (disabled during measurement or when the measurement switch is locked), with a warning beep before shutdown.

Clock settings: Year/month/day/hour/minute can be set, measurement data is automatically associated with timestamps, and there is a built-in lithium battery backup clock (with a lifespan of about 10 years).

Memory function: can store 1000 sets of measurement data (voltage/insulation resistance/low resistance), including DAR/PI, time, range and other information; Support data viewing and deletion (single/all), and transfer to PC through USB adapter.

image.png

System Settings and Data Management

1. Function configuration (long press SETUP to enter configuration mode)

Pass/Tail reference value: Set the judgment threshold for each voltage level, supporting preset values (such as default 0.5M Ω for 500V level) or custom values (0.001-4199M Ω).

Backlight/LED lights: Set to "ON" (automatic start stop) or "OFF" (normally off).

Buzzer: Set "ON" (live warning, discharge, on/off prompt) or "OFF" (turn off all beeps).

Clock calibration: Adjust the year/month/day/hour/minute to ensure accurate data timestamps.

2. Data management

Storage: When holding the measurement value, press the MemoRY button briefly to set the site number (SITE No.1/2, 0~99) and data number (automatically increasing) to complete the storage.

View: Press and hold the Memory button during standby time, select the data number using the directional keys, and view the measured values and station information.

Delete: When viewing data, select the number (or "ALL") and press ENTER to confirm the deletion (single or all).

Transmission: Connect to the PC via USB communication adapter (91030), install the driver, and export data using commands (refer to IM 91030-01EN manual).


Maintenance and repair

1. Daily maintenance

Battery management: When the battery level is low, the "battery icon" displays "empty" and needs to be replaced in a timely manner (4 AA alkaline batteries); Long term disuse requires removing the battery to avoid leaking and damaging the instrument.

Cleaning: Wipe the outer shell with a dry soft cloth, and do not use abrasives or solvents; The surface of the illuminance sensor needs to be kept clean to avoid affecting the automatic backlight/LED function.

Calibration cycle: It is recommended to calibrate once a year to ensure measurement accuracy. Calibration must be performed by Yokogawa certified personnel.

2. Fault handling and maintenance

Frequently Asked Questions:

Inaccurate measurement value: Check the battery level, probe connection, and perform 0 Ω calibration again for low resistance measurement.

Backlight/LED not on: Clean the illuminance sensor or manually confirm that the configuration mode is set to "ON".

Unable to store data: Memory is full (old data needs to be deleted), or the operation was not in the 'measurement value hold' state.

Maintenance restrictions: It is prohibited to disassemble the instrument by oneself (only battery replacement can open the battery compartment). For internal faults, please contact the Yokogawa dealer. Repairs involving safety insulation components require professional operation.

3. Disposal and Compliance

Equipment disposal: When disposing of equipment, local laws and regulations must be followed, and batteries (alkaline batteries and built-in lithium batteries) must be separated and classified for disposal.

Environmental Compliance: Compliant with EU RoHS Directive, WEEE Directive (prohibiting the mixing of household waste in EU regions), and EU Battery Directive (batteries must be recycled separately).


Product Specifications

Appearance dimensions: 156 (W) × 46 (H) × 97 (D) mm, weight approximately 490g (including battery).

Factory default settings: such as Pass/Tail reference values for each voltage level (0.5M Ω for 500V level), default "ON" for backlight/buzzer, default clearing of memory, etc.

  • EUCHNER HBA-079827 Pendant Station
  • CLC-2 4L PLC Lubrication Pump
  • KEYENCE GS-51P5 Safety Switch
  • AB 442G-MABH-R Safety Switch
  • GE Fanuc VersaMax PLC Module Set
  • Siemens 6ES7214-1HF40-0XB0 CPU 1214FC
  • Microchip DSPIC30F4011-30I/P DSC
  • FANUC A20B-2102-0081 I/O Link Module
  • Endress Hauser CLS15-B1M2A Conductivity Sensor
  • B&R 3AM050.6 Analog I/O Module
  • Fanuc A16B-2201-0320 MAIN-B CPU Board
  • Pilz 475650 PNOZ 1 Safety Gate Relay
  • Omron NSH5-AL001 Handheld HMI Terminal
  • Allen-Bradley 1756-OF8 Analog Output 8 Ch
  • Siemens 6SL3210-1SE31-0AA0 45kW Power Module
  • PMA TB45-110-00000-000 Temperature Limiter
  • PSR-SCP-24DC-ESD-5x1-1x2-300 Safety Relay
  • Pilz 774140 PZE 9 24V AC Safety Relay
  • Telemecanique TSXRKN82F 8 Slot Rack
  • Mitsubishi R16CPU iQ-R PLC CPU
  • Mitsubishi A2ACPU-R21-S1 PLC CPU
  • Omron NX-AD4208 Analog Input Unit
  • Schneider LMC802CAA10000 PacDrive 3 Controller
  • Reliance Electric 0-51874 Static Sequence Card
  • Pilz 787310 PNOZ X3P C Safety Relay
  • B&R X20CP1684 CPU Module
  • Siemens 6SN1145-1BB00-0FA1 Power Module
  • Beckhoff EL3174 Analog Input EtherCAT Terminal
  • CLC-2P 4L PLC Lubrication Pump System
  • Omron CJ1W-DA08C Analog Output Unit
  • Metso Automation D201776 ACN PO DC PLC Control Server Computer
  • GE AT868 AquaTrans Ultrasonic Flow Transmitter
  • ABB PFSA107-Z42 DTU Stressometer Digital Transmission Unit
  • ABB PFSA240 3BSE073476R1 Roll DC Supply Unit
  • Fanuc A16B-2201-0320 CPU MAIN Board
  • Pilz 475650 PNOZ 1 Safety Gate Relay
  • Omron NSH5-AL001 HMI Interface Unit
  • Allen-Bradley 1756-OF8 Analog Output Module
  • Siemens 6SL3210-1SE31-0AA0 Power Module 45kW
  • PMA TB45-110-00000-000 Temperature Limiter
  • PSR-SCP-24DC-ESD-5x1-1x2-300 Safety Relay
  • Pilz 774140 PZE 9 Safety Relay
  • Telemecanique TSXRKN82F PLC Rack Chassis
  • Mitsubishi R16CPU PLC CPU Module
  • OMRON C500-PS223-E Power Supply Module
  • Siemens 3VL4731-1DC36-0AA0 Circuit Breaker
  • Siemens 7ML5201-0EA0 Ultrasonic Level Transmitter
  • OMRON NQ3 NQ5 Touch Panel HMI
  • OMRON CJ1W-AD081-V1 Analog Input Module
  • OMRON NJ301-1100 Machine Automation Controller
  • B&R X20BC00G3 EtherCAT Bus Controller
  • Schneider ATV212HD22N4S Variable Speed Drive
  • B&R 8B0C0320HW00.002-1 Power Supply Module
  • Mitsubishi OSA105S2A Incremental Rotary Encoder
  • Pilz 777514 PNOZ XV3P Safety Relay
  • Gould AS-884A-111 Modicon 884 Controller
  • Siemens 6SC6130-0FE00 SIMODRIVE Control Card
  • Omron CV500-PS221 PLC Power Supply Module
  • ABB CM577-ETH AC500 PLC Ethernet Module
  • Omron NX-SIH400 Safety Input Unit NX Series
  • Omron NJ501-1300 Machine Automation Controller
  • Siemens 3VT8563-2AA03-2KA2 Molded Case Breaker
  • Pilz PNOZ m1p ETH 773103 Safety Controller
  • Omron CJ1H-CPU66H-R CJ1 Series CPU Module
  • ASI ASI533-S00 PLC Module S1
  • Mitsubishi AJ71C21-S1 Serial Module
  • Keyence IX-1000 Laser Sensor Amplifier
  • Siemens 6SN1145-1AA01-0AA1 Power Module
  • Siemens 3VA2340-5HL32-0AA0 MCCB 400A
  • Mitsubishi OSA104S Absolute Encoder
  • Siemens 6ES7350-1AH03-0AE0 FM 350-1 Counter
  • Siemens 6SE7038-6EK84-1JC2 IGD8 Gate Driver
  • Eaton EASY819-AC-RC Programmable Relay
  • Omron CPM1A-40CDT-D PLC 24V DC
  • Omron NA5-12W101B-V1 12-inch Programmable Terminal
  • Siemens 6ES7331-7KF02-0AB0 Analog Input SM 331
  • Moxa PTC-101-S-SC-HV Photoelectric Converter
  • Fanuc A20B-3300-0031 CNC Control Circuit Board
  • OMRON NA5-7W001B-V1 Programmable Terminal HMI
  • Parker AH385851U002 590C DC Drive Power Board
  • ABB 3BSE040662R1 AI830A Analog Input Module
  • DOLD BF9250.01/001 Solid State Relay
  • Siemens 6ES7331-7KF02-0AB0 SIMATIC S7-300 SM 331
  • ABB 07AC91 I6 GJR5252300R3101 Advant Controller 31
  • Schneider HMIST6500 Touch Screen HMI
  • Phoenix IL PN BK DI8 DO4 2TX-PAC Bus Coupler
  • NLTECH NL8060AC21-21D 8.4 inch LCD Display
  • Mitsubishi A2NCPU-P21 CPU Module A2NCPUP21
  • 1841-PL1-B-LH-ES Pressure Controller
  • Mitsubishi GT2512-STBA GT2512-STBD HMI
  • Cosel PJA600F-24 24V 600W Power Supply
  • B&R X67DM1321 Digital Mixed Module
  • Fanuc A16B-1310-0010 Power Supply Unit
  • Festo VABF-S4-1-V2B1-C-VH-20 Vacuum Generator
  • Mitsubishi A3ACPUP21 A3ACPU-P21 CPU
  • Mitsubishi Mazak PD14B-1 Power Supply
  • Siemens 6SY7000-0AC80 Module
  • PILZ 774400 PDZ Safety Relay
  • Mitsubishi A1SX81 Digital Input Module
  • Asyst Gasonics R94-1163 Controller 112671
  • Cincinnati Milacron 3 545 1000A Card
  • Siemens 6DD1607-0AA2 UPS Module
  • Square D 8536SCG3V02S Motor Starter
  • Siemens 6SL3100-1DE22-0AA1 Control Supply
  • Beckhoff EL9400 EtherCAT Power Terminal
  • Fanuc A16B-2202-0421 Power Supply Board
  • Turck TBEN-LG-8DIP-8DOP I/O Block
  • Euchner MGB-L1B-PNA-L-121861 Safety Switch
  • Pilz 772170 PNOZ M EF 1MM Safety Expansion
  • Mitsubishi AY51 MELSEC-A Output Module
  • Advantech AMAX-5056-A 8-Channel Digital Output
  • NXP MC912D60ACPV8 16-Bit MCU 60KB Flash
  • Omron C500-PS223-E SYSMAC PLC Power Supply Unit
  • Balluff BNI0047 BNI PBS-302-101-Z001 Profibus Module
  • Siemens 6SN1118-0DM31-0AA0 Regulation Card
  • Bently Nevada 9200-02-01-10-00 Sensor
  • Omron C500-PS223-E Power Supply Unit
  • Cognex COG-VC5-120-000 Vision Controller
  • ABB P-HC-BRC-40000000 Bridge Controller
  • Mitsubishi AJ71E71N-B5T PLC Module
  • Phoenix PSI-REP-PROFIBUS/12MB 2708863
  • Siemens 6SL3210-5BE31-8UV0 Converter
  • ABB CI868K01-eA 3BSE048845R2 Module
  • Honeywell S7999D1006 Touch Screen
  • Keyence CA-CHX10U Repeater Unit
  • Keyence VT5-W07 Touch Screen HMI
  • Keyence KV-8000 PLC CPU
  • Siemens 7MH4900-3AA01 Weighing Module