Welcome to the Industrial Automation website!

NameDescriptionContent
HONG  KANG
E-mail  
Password  
  
Forgot password?
  Register
当前位置:

TEKTRONIX TCPA300/400 current probe amplifier

来源: | 作者:FAN | 发布时间 :2025-11-21 | 170 次浏览: | Share:

TEKTRONIX TCPA300/400 current probe amplifier

Product positioning

Product portfolio:

Amplifier: TCPA300 (supports TCP312/305/303 probes), TCPA400 (only supports TCP404XL probes);

Probe: TCP312(30A/100MHz)、TCP305(50A/50MHz)、TCP303(150A/15MHz)、TCP404XL(750A/2MHz, Discontinuous).


Safe operation and system configuration

1. Core security standards

Prohibited live operation: Before connecting/disconnecting the probe/test line, the voltage source must be cut off to avoid insulation breakdown inside the probe;

Terminal matching requirements: The amplifier output needs to be connected to a 50 Ω load (oscilloscope or 50 Ω through terminal), otherwise the NOT TERMINATED IN 50 Ω LED will light up, affecting measurement accuracy;

Wire usage restrictions:

TCP312/305: Only insulated wires can be tested, and contact with exposed conductors is prohibited;

TCP303/404XL: can measure exposed conductors, but the probe needs to be removed and installed after power failure;

Environmental restrictions: Prohibit humid/explosive environments, working humidity of 5% -95% (≤ 30 ℃), 5% -85% (>30 ℃ -+50 ℃), non condensing.

2. System configuration and connection

(1) System composition

A complete current measurement system requires three parts:

TCPA300/400 amplifier: converts probe induced current into proportional voltage;

TCP300/400 series probe: clamp type induction current (effective after sliding cover locking);

Oscilloscope: requires a 50 Ω input impedance (if not available, add a 50 Ω straight through terminal), and the recommended bandwidth is 5 times the probe (e.g. TCP312 requires ≥ 500MHz oscilloscope).

(2) Connection steps

Amplifier → Oscilloscope:

TEKPROBE compatible oscilloscope: use TEKPROBE interface cable;

Non compatible oscilloscope: use 50 Ω BNC wire, set the oscilloscope input to 50 Ω;

Probe → Amplifier: Align the red marked points of the probe and amplifier, and lock them in place (do not twist);

Preheating: All equipment should be powered on and preheated for 20 minutes to ensure thermal stability.

image.png

Core Function Operation Guide

1. Key operations of the probe

(1) Demagnetization and automatic balancing (mandatory steps)

Function: Eliminate residual magnetism in the probe core, offset DC offset in the amplifier, and avoid measurement errors;

Trigger conditions: The probe slide is locked and disconnected from the live conductor;

Operation steps:

Press the PROBE DEGAUSS AUTOBALANCE button on the amplifier;

Wait for 5 seconds, the LED will flash from orange to green (successful), red (failed, terminal/probe connection needs to be checked);

Trigger scenarios: after startup, replacement of probe, overload, exposure to strong magnetic field.

(2) Sliding operation (different probe differences)

Probe model, sliding cover, opening method, locking method, precautions

TCP312/305 pulls the sliding cover backwards and pushes it forward until the buckle is fixed, only clamping the insulated wire

TCP303/404XL press lock+press handle release handle+press lock can clamp exposed wires (power-off operation)

2. Amplifier core control

(1) Key buttons and indicators

Control/indication function description exception handling

The COUPLING key switches between AC/DC coupling (AC only measures AC, DC measures AC/DC), and flashing indicates the display of Error Code

The RANGE key (TCPA300) switches the range (1/5/10/50A/V), and the LED indicates the current range. If the LED is not on, it means that the probe is not connected

PROBE OPEN LED probe slide not locked and turns off after locking the slide

OVERLOAD LED red: current beyond continuous range; Orange: Overtemperature; Flashing: Both exceed power-off cooling+demagnetization

NOT TERMINATED LED amplifier output not connected to 50 Ω load. Check oscilloscope impedance or add terminal

(2) Manual balancing (under DC coupling)

Function: Fine tune DC offset to improve measurement accuracy;

Operation: Press the MANUAL BALANCE key (left and right keys to adjust the offset direction), the LED lights up to indicate entering manual mode;

Limitation: Only DC coupling is effective, disabled under AC coupling.

3. Measurement process (taking DC current as an example)

Preheat the system for 20 minutes, set the amplifier to DC coupling, and set the oscilloscope input to 50 Ω;

Probe slide lock (no conductor), perform demagnetization/automatic balancing;

Adjust the grounding reference of the oscilloscope to the target scale;

Open the probe slide cover, clamp the tested conductor (probe arrow pointing in the normal current direction: positive → negative), and lock the slide cover;

Adjust the oscilloscope time base, trigger and gain, and read the current value (current=oscilloscope voltage value/amplifier range, for example, 2V corresponds to 10A at 5A/V range).

image.png

Performance validation and calibration

1. Preconditions for Performance Verification

Environmental conditions: temperature 23 ℃± 5 ℃, humidity meets requirements, no strong magnetic field;

Equipment requirements: current source (0.1% accuracy), DMM (5.5bit resolution), high-frequency current loop, etc. (see Table 5-3/5-11/5-17 for details);

Data recording: Corresponding test record forms (such as TCPA300 Table 5-16, TCP305 Table 5-29) need to be filled out.

2. Core testing items (amplifier+probe)

(1) DC gain accuracy test (taking TCP305 as an example)

Equipment connection: current source → 5-turn current loop → probe → amplifier → DMM (50 Ω terminal);

Test steps:

After demagnetization, the probe clamps the current loop and the current source outputs+1.00A;

Record the DMM voltage value M1, and then output -1.00A to record M2;

Calculation error:% Error=[(M1-M2) -2 × VE]/(2 × VE) × 100% (VE is the expected voltage, such as 1A corresponding to 1V in the 5A/V range);

Qualified range: ≤± 3% (guaranteed value), ≤± 1% (typical value).

(2) Bandwidth test (using TCP312 as an example)

Equipment connection: level sine wave generator → high-frequency current loop → probe → amplifier → oscilloscope;

Test steps:

The generator outputs a 3MHz, 3Vpp signal and records the oscilloscope peak to peak value M1;

Adjust the generator frequency to 100MHz (TCP312 bandwidth) and record M2;

Verification: (M2/M1) × 1.18 ≥ 0.707 (1.18 is the impedance correction factor);

Qualified range: -3dB bandwidth ≥ 100MHz.

(3) Rise time test (taking TCP404XL as an example)

Equipment connection: High voltage pulse generator → Built in current loop → Probe → Amplifier → Oscilloscope;

Test steps:

The generator outputs 5A, 250 μ s pulses, and the oscilloscope is set to 80ns/div;

Measure the rise time of pulses ranging from 10% to 90%;

Qualified range: ≤ 175ns (derived from a bandwidth of 2MHz: 0.35/2MHz=175ns).

3. Calibration adjustment (only operated by qualified personnel)

(1) Amplifier adjustment (TCPA300)

Prerequisite: The casing needs to be disassembled and a calibration adapter (174-4765-00) needs to be used;

Step: The current source outputs 0.100A, adjust the corresponding range potentiometer (such as adjusting the 25 Ω potentiometer for 5A/V) to make the DMM voltage meet the expected value (such as 0.1A corresponding to 0.25V at 5A/V range).

(2) Probe adjustment (TCP305/312)

DC gain adjustment: Open the probe housing and adjust the DC gain potentiometer to make the DMM voltage error ≤ ± 3% (refer to Table 6-7);

Attention: Record the original position before adjustment to avoid exceeding the specifications.

Troubleshooting and Maintenance

1. Common faults and solutions

Possible causes and solutions for the fault phenomenon

Demagnetization failed (LED red), 50 Ω load not connected, probe not locked, check terminal+lock slide cover

OVERLOAD LED constantly on, current exceeding range, probe overheating, current reduction+power-off cooling+demagnetization

The measurement value is not accurate without demagnetization, the range is selected incorrectly, the terminal is not connected to demagnetization+check the range+check the 50 Ω load

The probe without output signal is not connected, and the coupling is set to AC (measuring DC). Connect the probe and switch the DC coupling

2. Interpretation of Error Code

Display mode: AC/DC coupled LED flashes alternately, and 4 fault LEDs (PROBE OPEN is MSB, NONCOMPATIBLE is LSB) display a 4-digit binary code;

Common code:

Code 1: No Hall element detected in the probe → Reconnect or replace the probe;

Code 2: DC offset calibration failed → restart amplifier+demagnetization;

Code 15: Internal software error - If restarting is ineffective, contact after-sales service.

3. Daily maintenance

Cleaning: Wipe the outer shell with 75% isopropanol, and do not use solvents such as benzene/acetone;

Storage: Non working temperature -40 ℃ -+75 ℃, avoid moisture/dust;

Calibration cycle: It is recommended to calibrate once a year, with a reference frequency error (± 10ppm) valid for one year.


Accessories and Replacement Parts

1. Standard accessories (included with the product)

Category, accessory name, purpose

Amplifier accessory 50 Ω straight through terminal (011-0049-02) matching output impedance

Connect the TEKPROBE interface cable (012-1605-00) to the TEKPROBE oscilloscope

BNC line (012-0117-00) connected to non TEKPROBE oscilloscope

Probe accessories Probe cover (016-1923-00) protects the probe

Grounding lead (196-3120-01) probe grounding

2. Key replacement parts (partial)

Component type, model, and part number

TCPA300 motherboard circuit motherboard 671-5434-00

TCP305 coil current transformer component 120-184-00

TCP404XL cable probe cable assembly 174-4816-00

Amplifier fan axial fan (12VDC) 119-6721-00

image.png

  • WESTINGHOUSE 1C31113G02 Ovation Analog Input Module
  • WESTINGHOUSE 1C31129G03 Control Module
  • WESTINGHOUSE 1C31122G01 Process Controller | Ovation DCS Control Module
  • WESTINGHOUSE 1C31113G02 Ovation Analog Input Module
  • WESTINGHOUSE 1B30023H02 Control Module
  • WESTINGHOUSE 1B30035H01 Turbine Control System Module
  • WIDAP UFW30.156 6K8J175W0823 Power Resistor Technical Profile
  • WINGREEN IPB PCB V2.0_A01 03ZSTL6-00-201-RS Industrial Power Board
  • WINGREEN CANopen_ADAPTER V5.0_A01 03ZSTI-00-501-RS Module
  • WINGREEN PUIM V2.0 034STM4-00-200-RS Power Interface Module
  • WINGREEN DUDT_DETECTION_V2.0_A01 03ZSTJ0-00-201-RS Detection Control Board
  • WINGREEN LAIB V3.0_A00 034STN1-00-300-RS Embedded Industrial Motherboard
  • WINGREEN FAN_DETECTION V1.0_A05 03ZSTJ3-00-105Fan Monitoring Module
  • WINGREEN LAIB V3.0_A00 034STN1-01-300-RS Interface Board
  • WINGREEN ATKB_V5.0_A01 03ZSTI4-00-501 Industrial Control Keyboard Module
  • WINGREEN ATKB_V5.0_A01 03ZSTI4-01-501 Industrial Motherboard | Embedded Control Board
  • WINGREEN FPB_V3.0_A01 03ZSTJ1-00-301-RS Fieldbus Processor
  • WINGREEN DSPB_V4.0_A02 03ZSTI7-00-402-RS Digital Processing Board
  • WOHNER 31110 Cylindrical Fuse Holder
  • WOODHEAD APPLICOM PCI4000 PCI Communication Card Industrial DeviceNet CAN Bus Interface
  • Woodward 8440-1706 Industrial Control System Module
  • Woodward 8440-2052 H Synchronizer Load Share Module
  • Baldor KPD-TS12C-30E 12.1" Color TFT Touch Screen Ethernet HMI
  • Baldor KPD-TS10C-30E 10" Color TFT Touch Screen Operator Interface with Serial and Ethernet Interfaces
  • Baldor KPD-TS05C-30E 5.6" Color TFT Touch Screen with Serial and Ethernet Interface
  • Baldor KPD-TS05C-30 5.6 Inch Color TFT Touch Screen Serial Interface
  • Baldor KPD-TS05M-10 5.6" Monochrome Touch Screen Serial Interface HMI
  • Baldor KPD-TS03M-10 Monochrome Touch Screen Operator Interface
  • Baldor KPD-KG420-30 4x20 Graphic Display with 12 Function Keys - Serial Interface
  • Baldor KPD-KG420-20 4x20 Character Graphic Display Serial Interface
  • WOODWARD EASYGEN-3200-5 8440-1992 A Genset Controller
  • WOODWARD PEAK200-HVAC 8200-1501 C Version | Industrial Building Automation Controller
  • Woodward 8440-2052 easyGEN-3200 Genset Control Power Management
  • Woodward 8237-1246 + 5437-1119 Control System Module
  • WOODWARD SPM-D11 8440-1703 Overspeed Protection System Module
  • WOODWARD 8237-1369 Governor Control Module
  • Woodward 8237-1600 Digital Control Module
  • WOODWARD BUM60-1224-54-B-001-VC-A0-0093-0013-G003-0000 3522-1004 Industrial Control Module
  • WOODWARD 8200-1302 Genset Controller
  • Woodward 8901-457 Speed Control Module
  • WOODWARD 5501-465 Control Module
  • Woodward 5448-890 SPM-D10 Digital Control Module
  • WOODWARD 5437-1067A Turbine Governor Actuator
  • Woodward 8440-1666 B Digital Control Module
  • WOODWARD 8440-1706 A SPM-D11 Synchronous Phase Modulator Module
  • WOODWARD 5466-425 Programmable Automation Controller (PAC)
  • WOODWARD 5466-318- Industrial Gas Turbine Control Module
  • WOODWARD 5453-277 Digital Control Module
  • WOODWARD 5453-203 Digital Governor Control Module
  • WOODWARD 9907-1106 Pressure Converter
  • WOODWARD 5233-2089 Professional Industrial Control System Module
  • WOODWARD 9907-147 Power outage tripping overspeed protection system
  • WOODWARD 8237-1600 Digital Speed Control System
  • WOODWARD 8402-319 8402-119 microprocessor speed controller
  • Woodward 8237-1006 Digital Governor
  • WOODWARD 5501-471 Communication Module
  • WOODWARD 5466-258 Input/Output Module
  • WOODWARD 5501-467 Multi Protocol Communication Gateway and I/O Expansion Module
  • WOODWARD 5501-470 Digital microprocessor controller module
  • WOODWARD 9907-1200 Digital Governor
  • WOODWARD 8444-1067 High Performance Digital Microprocessor Controller Module
  • WOODWARD 8446-1019 Integrated Gas Engine Electronic Control System
  • WOODWARD 9907-162 Digital Engine Governor
  • WOODWARD 5466-316 Simulation Combination Module
  • WOODWARD 5464-414 Digital Speaker Sensor Module
  • XANTREX XFR40-70 DC power supply
  • XP POWER F8B6A4A6A6 power module
  • XP POWER F8B6D4A3G3 power supply
  • XYCOM XVME-674 VMEbus Single Slot CPU/Processor Module
  • XYCOM XVME-957 Circuit Board
  • XYCOM XVME-976 PC board computer
  • XYCOM XVME-530 8-Channel Isolated Analog Output Module
  • XYCOM Proto XVME-085 Bus Module
  • YAMAHA RCX40 4-AXIS ROBOT CONTROLLER
  • YAMATAKE EST0240Z05WBX00 touch screen display
  • YAMATAKE HD-CAOBS00 flowmeter
  • HIMA X-COM 01 Communication Module
  • HIMA HIMax X-AO 16 01 Analog Output Module
  • HIMA X-AI3251 Analog Input Module
  • HIMA X-DO3251 Digital Output Module
  • HIMA X-DI3202 Digital Input Module
  • HIMA X-DI6451 Digital Input Module
  • YASKAWA USAHEM-02-TE53 AC servo motor
  • Yaskawa JZNC-XPP02B Teaching Programmer
  • YASKAWA CACR-SR07BE12M servo drive
  • YASKAWA JAMSC-B2732V Advanced Drive Controller
  • YASKAWA JGSM-06 Controller
  • YASKAWA PCCF-H64MS 64MB Industrial Memory Module
  • YASKAWA CACR-02-TE1K servo driver
  • YASKAWA JAPMC-IQ2303 Controller Module
  • YASKAWA DDSCR-R84H Controller
  • YASKAWA JANCD-XTU01B circuit board
  • YASKAWA JANCD-XIO01 High Performance PC Input/Output (I/O) Board
  • YASKAWA JACP-317800 servo drive
  • XYCOM 120974 - Circuit Board
  • XYCOM 99298-200 - PC Control Card 99207A-001
  • XYCOM 99298-266 - CPU Board
  • XYCOM 99311-001 - Screen Display Ribbon Cable
  • XYCOM 99384-002 - PCB Module 99383A-001
  • XYCOM A05520I01 - Control Module PW22471 MPL400
  • XYCOM A33369 - Dual PMC Carrier Module
  • XYCOM AC2064A1 - VMEbus PCB Board Analog Output
  • XYCOM AC2065A1 - VMEbus PCB Board Analog Output
  • XYCOM CDA XVME - DSP SCSI I/O Module Serial RS232
  • XYCOM 143913-001 - TSKAM Board Rev A
  • XYCOM 1300-000100000 - Operator Interface Node Module
  • XYCOM 140050(R) - Documentation Support Library
  • XYCOM 1811-A00000000 - SXT1811T Monitor 100-120V
  • XYCOM 3112T-1200-256-2K - Operator Interface
  • XYCOM 3115 T - Industrial Computer
  • XYCOM 3406T - Flat Panel Industrial PC
  • XYCOM 3512-03F114003 - Automation Operator Interface
  • XYCOM 3515-A2P214003 - Operators Interface 6.3A
  • XYCOM 3535 - Operation Panel 3535-E3T434003
  • XYCOM 3612 T - LCD Flat Panel Display
  • XYCOM 3715 - Industrial Touchpanel Touchscreen PC
  • XYCOM 4115 T - Light-Duty Flat Panel 2000-512-2K
  • XYCOM 50151 - Automation Display
  • XYCOM 5017-T - Flat Panel Industrial Touch Monitor
  • XYCOM 3308T - 8" Flat Panel Color Touchscreen PC
  • XYCOM 9000-FFP-HU - Industrial Module
  • XYCOM 9440 - Automation HMI 9440-026714001
  • XYCOM 3406 KPT - Automation Control Panel
  • XYCOM 3000-CBL-25VP - Video Cable 25' 7.5M
  • XYCOM CA3-MBPALL-41 - Modbus Plus Module for GP2000
  • XYCOM QPI-MBP-XA1 - Modbus Plus Communication Module
  • XYCOM 3512 T - Automation Model HMI PM-101722C
  • XYCOM PM-102159 - Xycom Automation Monitor