Basic Information and Quality Assurance Policy
Product positioning and applicable scenarios
Core function: Measure the static/dynamic characteristics of semiconductor devices and support the drawing of voltage current (V-I) characteristic curves for bipolar transistors, field-effect transistors (FETs), diodes, thyristors (SCR), and other devices.
Key advantages: digital storage (flicker free display), precise cursor measurement (minimum resolution of 100pA), GPIB remote control, pulse mode (reducing device power consumption).
Safety regulations and system installation
1. Core security requirements
Interlocking system: The test adapter must be equipped with a protective cover. When there is no cover, the collector power supply will be automatically disabled, and the red Warning light will turn off (the light will turn on when there is a cover, indicating dangerous voltage).
Live operation taboos: Do not plug or unplug when connecting the probe/test line to the voltage source, and do not remove the protective cover for measurement.
Power safety: Before replacing the power supply, it is necessary to cut off the power and confirm that the LINE VOLTAGE SELECTOR (115V/230V) matches the local power supply. The fuse specifications cannot be mixed (4A for 115V and 2A for 230V).
2. System installation process
(1) Power configuration
Confirm the LINE VOLTAGE SELECTOR on the Rear panel: select 115V/230V for NOMINAL and HIGH-LOW for RANGE (e.g. 115V HIGH corresponds to 107-132V).
Check fuses: Open the Rear panel fuse holder. For 115V models, use a 125V/4A fast melting fuse (159-0259-00), and for 230V models, use a 250V/2A fast melting fuse (159-0260-00).
Connect the power cord: Select the corresponding plug according to the region (such as standard plug for North America or A1 option plug for Europe).
(2) GPIB settings
Rear panel GPIB Terminator and Address switch: Set the address (0-31, 31 is offline) and select the termination symbol (EOI/LF).
Cable connection: Use standard GPIB cables (such as 012-0991-00, 2m) to connect the controller to 370B, with a maximum of 15 devices in the system.
Address verification: Press and hold the FAST/SHIFT+LOCAL keys, and the CRT will display the GPIB address and termination symbol.
(3) Test adapter installation
Standard adapters: A1001 (blank), A1002 (serial), A1005 (axial lead), align with A1006-A1010 interface during installation, and do not forcefully install double width adapters to the right interface.

Core Function Operation Guide
1. Key controls on the front-end panel
(1) Collector Supply Control
Key parameters of control item function description
MAX PEAK VOLTS selects peak voltage range 16V/80V/400V/2000V (4th gear)
MAX PEAK POWER selects maximum peak power of 0.08W/0.4W/2W/10W/50W/220W (6th gear)
Polarity selects 7 output polarities and modes:+LEAKAGE (measuring emitter current, sensitivity x 1000),+DC, AC, etc
VARIABLE regulates the output ratio from 0.0% to 100.0% (with a step size of 0.1%)
LOOPING COMPENSATION Compensation parasitic capacitance maximum compensation 100pF
(2) Step Generator Control
Key parameters of control item function description
OUTPUT MODE Select output type: current (50nA-200mA/step), voltage (50mV-2V/step)
NUMBER OF STEPS: Select the number of steps from 0 to 10 (only output offset when 0 steps)
PULSE selects the pulse modes SHORT (80 μ s), LONG (300 μ s), OFF
OFFSET adjustment offset ± 10 x step amplitude (step size 1%)
Inverted output polarity is only valid in EMITTER COMMON mode
(3) Digital Storage and Display Control
Storage mode:
NON STORE: Real time simulation display (no storage);
STORE: Digital storage (1024 points/axis, resolution 10 bits);
VIEW: Call the storage curve (16 IC memories/64 floppy disks).
Collection mode:
AVG: 16 times average (noise reduction);
ENV: Vertical/Horizontal Envelope (Thermal Drift);
Norm: Conventional collection.
Cursor measurement:
DOT: Single point measurement (reading voltage/current);
F LINE: Functional line measurement (on resistance, intercept voltage);
WINDOW: Window measurement (hFE, gm).
2. Basic measurement process (taking diode forward characteristics as an example)
Install adapter: Insert the A1005 axial lead adapter into the right interface, insert the diode (anode connected to collector, cathode connected to emitter), and close the protective cover.
Configure collector power supply:
Choose 16V for MAX PEAK VOLTS, 0.4W for MAX PEAK POWER, and+DC for POLYTY.
Configure stepper generator:
Select CURRENT for OUTPUT MODE, 1mA/step for STEP AMPLITUDE, and 5 for NUMBER OF STEPS.
Start measurement:
Press MEASURE → REPEAT, adjust the VARIABLE collector power supply to 100%, and the CRT displays a positive V-I curve.
Cursor measurement:
Press CURSOR → DOT, move the cursor to the conductive area, and the CRT displays the forward voltage (such as 0.7V) and current (such as 5mA).

Typical measurement applications
1. Bipolar transistor (NPN type)
Measurement parameters: IC-VCE curve hFE、VCE (sat)、ICE0;
Key settings:
Configuration: CONFIG → BASE STEP GEN (emitter common ground);
Collector power supply: MAX PEAK VOLTS=80V, POLYITY=+DC;
Stepper generator: CURRENT mode, 50 μ A/step, 10 steps;