Welcome to the Industrial Automation website!

NameDescriptionContent
XING-Automation
E-mail  
Password  
  
Forgot password?
  Register
当前位置:

YOKOGAWA ZR22G, ZR402G split zirconia oxygen/humidity analyzer

F: | Au:FAN | DA:2025-12-18 | 936 Br: | 🔊 点击朗读正文 ❚❚ ▶ | Share:

Detector connection: including sensor signal (CELL ±), thermocouple (TC ±), cold junction compensation (CJ ±), heater power supply (HTR) wiring.

External interfaces: analog output (4-20mA), contact input/output (control calibration, alarm, etc.), HART communication (refer to dedicated documentation).

image.png

Operation and Calibration Process

(1) Startup process

Pre inspection: Confirm that there is no leakage in the pipeline, the wiring is correct, and the valve status is normal.

Power supply startup: Connect the converter power supply, and the sensor begins to preheat (about 20 minutes, target temperature 750 ℃).

Parameter confirmation: Check the initial parameters such as converter type (oxygen/humidity), detector type, sample gas type (wet/dry gas), output range, etc.

Output inspection: Confirm that the current circuit (4-20mA) and contact I/O function are normal.

(2) Calibration operation (core process)

1. Calibration preparation

Gas requirements: Zero gas (0.95-1.0 vol% O ₂+N ₂), span gas (clean and dry air, dew point ≤ -20 ℃).

Flow setting: Calibrate air flow rate of 600 ± 60 ml/min, reference air flow rate of 800-1000 ml/min.

2. Calibration mode

Applicable scenarios for mode operation mode

Manual calibration is performed by manually controlling the gas valve through the ZO21S/ZA8F unit, and completing the calibration system 1/2 according to the touch screen prompts. Manual intervention is required

Semi automatic calibration of touch panel or touch input startup, completing calibration system 3 according to preset time, reducing manual operation

Automatic calibration is automatically executed at set intervals without the need for manual intervention. System 3 operates continuously for a long time

3. Core calibration steps

Span calibration: Introduce span gas and confirm calibration after the value stabilizes.

Zero calibration: Introduce zero gas and complete the calibration in the same way.

After calibration: Close the calibration gas valve and wait for the output to stabilize (default 10 minutes).

(3) Key parameter settings

1. Output settings

Oxygen analyzer: Set oxygen concentration values corresponding to 4mA (minimum range) and 20mA (maximum range), supporting linear or logarithmic output modes.

Humidity analyzer: Set the humidity (volume fraction or mixing ratio) corresponding to 4mA/20mA, and input the measured gas temperature and pressure to calculate the relative humidity and dew point.

2. Alarm setting

Supports four levels of alarms: high high, high, low, and low low. Alarm thresholds, hysteresis values (0-9.9 vol% O ₂, etc.), and delay times (0-255 seconds) can be set.

Alarm output: 4 contact outputs that can be assigned to alarm, fault, calibration, and other states.


Maintenance and troubleshooting

(1) Regular maintenance

1. Daily maintenance (recommended cycle)

Cleaning: Calibration gas pipeline (anti clogging), detector dust filter (K9471UA).

Inspection: Pipeline leakage, valve status, and cable connection tightness.

2. Regularly replace components

Precautions for component replacement

The zero gas calibration ratio of the sensor component exceeds 100 ± 30%, and the span gas calibration ratio exceeds 0 ± 18%; When the internal resistance of cells significantly increases, the metal O-ring (K9470BJ) and contact (E7042BS) must be replaced simultaneously

Abnormal resistance of heater unit (≤ 90 Ω is normal), unstable heating temperature at 750 ℃, please refer to the dedicated heater manual (IM 11M12A01-21E)

Before replacing the fuse converter with a blown fuse (specification 3.15A, Part No.A1113EF), it is necessary to investigate the short circuit fault

(2) Common faults and solutions

1. Error fault (Error code)

Key points for troubleshooting the causes of error codes and malfunctions

Error 1 Cell voltage fault (<-50mV) Check CELL ± wiring, sensor damage, internal wiring of detector

Error 2: Heater temperature fault (below 730 ℃ or above 780 ℃). Check the heater resistance, thermocouple wiring polarity, and cold end temperature (-25-155 ℃)

Error 3: The A/D converter has malfunctioned and restarted the device. If the error persists, please contact after-sales service

Error 4: Memory failure. Restart the device, but the error still persists. Contact after-sales service for assistance

2. Alarm fault (Alarm code)

Key points for troubleshooting the causes of alarm code faults

Alarm 6 zero calibration coefficient abnormality (exceeding 100 ± 30%) confirms zero gas concentration, calibration gas flow rate, and sensor contamination situation

Abnormal calibration coefficient of Alarm 7 span (exceeding 0 ± 18%) confirms span gas concentration, pipeline leakage, and sensor aging

Alarm 8 calibration stability timeout check calibration air flow rate, pipeline length/diameter, sensor response speed

Alarm 10 Cold End Temperature Abnormal (< -25 ℃ or > 155 ℃) Check the wiring of the cold end sensor and the ambient temperature of the detector

3. Abnormal measurement values

High measurement value (oxygen analyzer): reference gas humidity changes, calibration gas leakage, abnormal increase in sample pressure.

Low measurement value (oxygen analyzer): zero gas leakage, combustible components in sample gas, sensor damage.

  • National Instruments PXI-5421 Arbitrary Waveform Generator
  • National Instruments cRIO-9023 CompactRIO Controller
  • National Instruments PCI-6221 Multifunction DAQ Card
  • National Instruments PXI-4065 Digital Multimeter Card
  • National Instruments NI 9477 Digital Output Module
  • National Instruments cDAQ-9185 TSN Ethernet Chassis
  • National Instruments NI 9147 4-Slot cRIO Chassis
  • National Instruments USB-6259 Mass Termination DAQ Device
  • National Instruments 184120F-01 Ordering Part Number
  • National Instruments PCI-4472B Sound and Vibration Module
  • National Instruments SCXI-1130 Switch
  • National Instruments VB-8012 VirtualBench
  • National Instruments PXI-7350 Controller
  • National Instruments eDAQ-9174 Chassis
  • National Instruments PXI-4461 DSA Module
  • National Instruments PCI-6280 DAQ Card
  • National Instruments PXI-6251 DAQ Card
  • National Instruments PXI-6653 Module
  • National Instruments PCI-6509 DIO Device
  • National Instruments NI 9751 Driver Module
  • National Instruments PXIe-6361 Multifunction I/O Module
  • National Instruments cRIO-9014 CompactRIO Controller
  • National Instruments PCI-7390 Motion Control Card
  • National Instruments PXI-8512 CAN Interface Module
  • National Instruments sbRIO-9636 Single-Board Controller
  • National Instruments IC-3120 Industrial Controller
  • National Instruments PCI-6220 M Series DAQ Card
  • National Instruments PXI-8176 Embedded Controller
  • National Instruments PXI-6132 S Series DAQ Module
  • National Instruments NI 9206 Analog Input Module
  • National Instruments PXI-2584 Switch Module
  • National Instruments 196541B-01L Module
  • National Instruments 778075-01 Module
  • National Instruments FP-RLY-420 Module
  • National Instruments USRP-2920 Radio
  • National Instruments PCIe-6351 DAQ Card
  • National Instruments cRIO-9040 Controller
  • National Instruments NI 9201 Input Module
  • National Instruments NI 9411 Input Module
  • National Instruments NI 9216 Input Module
  • National Instruments PCI-6221 DAQ Device
  • National Instruments PXI-4498 DSA Module
  • National Instruments PXI-6259 DAQ Card
  • National Instruments PXI-7354 Motion Controller
  • National Instruments FP-AI-110 Input Module
  • National Instruments 192172F-04L Module
  • National Instruments PXIe-8115 Controller
  • National Instruments 142931B-01L Module
  • National Instruments PXI-2503 Switch Module
  • National Instruments 776246-01 Module
  • National Instruments CFP-2110 Compact FieldPoint Controller
  • National Instruments 779429-01 Ordering Part Number
  • National Instruments NI-9149 8-Slot EtherCAT Chassis
  • National Instruments PXI-4110 Programmable DC Supply
  • National Instruments 192061-10 Ordering Part Number
  • National Instruments cDAQ-9181 Chassis Module Carrier
  • National Instruments NI 9437 Digital Input Module
  • National Instruments cFP-2020 Compact FieldPoint Controller
  • National Instruments PCI-6259 Multifunction DAQ Card
  • National Instruments FP-TC-120 Thermocouple Module
  • National Instruments NI-9174 CompactDAQ Chassis
  • National Instruments PXI-6030E DAQ Module
  • National Instruments PXIe-5160 Oscilloscope
  • National Instruments PXI-4462 DSA Module
  • National Instruments PXI-8109 Controller
  • National Instruments PXI-6259 DAQ Card
  • National Instruments PXI-6030E DAQ Card
  • National Instruments cRIO-9056 Controller
  • National Instruments PXIe-8370 MXI Kit
  • National Instruments USB-6212 DAQ Device
  • National Instruments RMX-4000 Electronic Load System
  • National Instruments 785040-01 Ordering Part Number
  • National Instruments cFP-AI-110 Analog Input Module
  • National Instruments cDAQ-9189 8-Slot Chassis
  • National Instruments PXIe-4139 Source Measure Unit
  • National Instruments PXI-5620 Digitizer Module
  • National Instruments PCIe-8371 MXI-Express Card
  • National Instruments 196762B-01L Ordering Part Number
  • National Instruments NI-9361 Counter Input Module
  • National Instruments cDAQ-9181 Ethernet Chassis
  • National Instruments NI 9221 Input Module
  • National Instruments cDAQ-9171 Chassis
  • National Instruments NI 9225 Input Module
  • National Instruments NI 9171 Chassis
  • National Instruments 779373-01 Module
  • National Instruments PCI-6034E DAQ Card
  • National Instruments NI-9211 Input Module
  • National Instruments 779013-01 Module
  • National Instruments NI-9159 RIO Chassis
  • National Instruments PCIe-GPIB Analyzer Card 778930-01
  • National Instruments PCI-6110 Multifunction DAQ Card
  • National Instruments PXIe-8840QC Embedded Controller
  • National Instruments PXI-2530 Switch Module
  • National Instruments 780927-01 Ordering Part Number
  • National Instruments PXI-8420 Serial Interface Module
  • National Instruments PXIe-8135 Embedded Controller
  • National Instruments GPIB-ENET/1000 Controller
  • National Instruments PCI-6704 Analog Output Card
  • National Instruments PXI-6133 S Series DAQ Module
  • National Instruments PXI-1042 8-Slot Chassis
  • National Instruments NI 9512 Module
  • National Instruments 779003-01 Module
  • National Instruments cDAQ-9191 Chassis
  • National Instruments USRP-2921 Radio
  • National Instruments 783368-01 Module
  • National Instruments PXI-2510 Switch Module
  • National Instruments PCI-7344 Controller
  • National Instruments PXIe-1082 Chassis
  • National Instruments PXI-8108 Controller
  • National Instruments USB-6363 DAQ Device
  • National Instruments cDAQ-9191 Wireless Chassis
  • National Instruments PXIe-4357 RTD Input Module
  • National Instruments PCI-6512 Industrial Digital I/O Card
  • National Instruments PXI-6115 Multifunction DAQ Module
  • National Instruments NI-9231 Sound and Vibration Module
  • National Instruments PCI-7344 Motion Controller
  • National Instruments PCI-8512 CAN Interface Device
  • National Instruments PCI-1409 Image Acquisition Card
  • National Instruments 783451-01 Ordering Part Number
  • National Instruments 781496-01 cDAQ-9181 Chassis
  • National Instruments PCI-6250 Multifunction DAQ Device
  • National Instruments PCI-6224 Multifunction DAQ Card
  • National Instruments cRIO-9081 CompactRIO Controller
  • National Instruments USB-6353 X Series DAQ Device
  • National Instruments NI-9223 Simultaneous Analog Input Module
  • National Instruments PXIe-8133 Embedded Controller
  • National Instruments NI-9403 Digital I/O Module
  • National Instruments PCI-6111E Multifunction I/O Card