Resolution 12 bits/0.2 ° C (i.e. minimum temperature detection accuracy of 0.2 ℃) 12 bits/0.5 ° C (i.e. minimum temperature detection accuracy of 0.5 ℃)
Precautions for replacing fuses
When the P (External Power) LED goes out and the fuse is suspected to be blown, the following steps must be followed to replace it, and safety regulations must be followed:
Turn off the PLC power first, and then remove the TCM module;
Use a small standard screwdriver (pocket size) to loosen the side card of the front cover and gently pull the cover outward;
Use pointed nose pliers to pull out the old fuse from the front of the module and replace it with a fuse of the same specification (2A, 125V);
Warning: It is strictly prohibited to use fuses that do not meet the specifications, otherwise it may cause personal injury or equipment damage.