2. Collection and analysis functions
(1) Sequence collection
Function definition: Divide long signals into multiple "sequence segments" for acquisition, with each segment recording a length that can be set (such as 1Mpts), and a maximum of 1000 segments in total;
Advantages: Avoiding excessive memory usage for a single long record and capturing burst signals with long intervals (such as 1-second pulse sequences).
(2) Signal analysis
FFT analysis: Supports 1024-1M point FFT, frequency range 0Hz Nyquist frequency, and provides 5 window functions including Hanning/Flattop/Rectangle;
Template testing: Customize waveform templates (such as standard sine wave contours), the oscilloscope automatically compares the collected waveform with the template, marks out out of tolerance areas, and is suitable for batch signal quality testing;
Measurement function: Supports 56 types of automatic measurements (frequency, cycle, peak to peak, rise time, etc.), measurement statistics (average, maximum, minimum, standard deviation), and supports user-defined measurement formulas.
3. Serial bus decoding and triggering
Supports bus types: covering 18 commonly used serial buses, including:
Low speed bus: I2C, SPI, UART (RS-232/422/485), LIN, CAN/CAN FD;
High speed bus: Ethernet (10/100/1000BASE-T), USB 2.0, SATA, PCIe (Gen1/2);
Decoding function: Real time decoding of bus data, displayed in color tags (such as I2C address in yellow and data in blue), supporting viewing of "event table" (displaying timestamp, type, and value of each frame of data);
Trigger function: It can be triggered based on specific bus conditions (such as I2C write address 0x50, CAN ID=0x123) to accurately capture target data frames.
4. Expansion and connectivity capabilities
Built in expansion module:
Function generator: 25MHz output, supporting 14 waveforms including sine wave (0.1Hz-25MHz), square wave (1Hz-25MHz, duty cycle 1% -99%), triangular wave/sawtooth wave (0.1Hz-1MHz), etc;
Logic analyzer: The MSO model has built-in 16 channel logic analysis function, which does not require additional hardware;
Protocol analyzer: supports bus data storage and playback, and can be exported in CSV/Excel format.
External interface:
Data transmission: USB 3.0 (device side, high-speed transmission of waveform data), USB 2.0 (host side, connected to USB flash drive/mouse), Ethernet (10/100/1000BASE-T, supporting LXI-C standard, remote control);
Display output: HDMI (1080P, external monitor);
Synchronization interface: Trigger In/Out (trigger signal input/output), Ext Clock (external clock input);
Storage capacity: Built in 16GB SSD (for storing waveforms/settings/decoding data), supports external USB drives (up to 2TB), and portable hard drives.
Comparison of Parameters for All Series Models (Core Models)
Model Channel Type Bandwidth Sampling Rate (Full Channel/Half Channel) Digital Channel Function Generator Typical Application Scenarios
DPO5104B 4 simulates 100MHz 2.5GS/s/5GS/s without optional basic analog signal debugging (such as power supply)
MSO5104B 4 Analog+16 Digital 100MHz 2.5GS/s/5GS/s 16 Optional Embedded System Mixed Signal Debugging
DPO5154B 4 Analog 150MHz 2.5GS/s/5GS/s No selectable medium speed analog signal (such as 100Mbps communication)
MSO5154B 4 analog+16 digital 150MHz 2.5GS/s/5GS/s 16 optional medium speed mixed signal (such as CAN bus)
DPO5504B 4 Analog 500MHz 2.5GS/s/5GS/s No selectable high-speed analog signal (such as 500MHz clock)
MSO5504B 4 analog+16 digital 500MHz 2.5GS/s/5GS/s 16 optional high-speed mixed signal (such as SPI high-speed communication)
DPO5204B 4 analog 2GHz 2.5GS/s/5GS/s no optional ultra high speed analog signal (such as 1GHz RF)
MSO5204B 4 Analog+16 Digital 2GHz 2.5GS/s/5GS/s 16 Optional Ultra High Speed Mixed Signal (such as PCIe)
DPO5104B-L 4 simulation (low noise) 100MHz 2.5GS/s/5GS No selectable low noise scenarios (such as sensor signals)
Compliance certification
Safety certification: compliant with UL 61010-1, CSA C22.2 No. 61010-1, IEC 61010-1 (overvoltage category II, pollution level 2);
EMC standards: comply with EN 61326-1 (industrial environment), CISPR 11 (Class A, non residential area), FCC Part 15 (Class A);
Environmental requirements: working temperature 0-50 ℃, storage temperature -40-70 ℃, relative humidity 5% -90% (no condensation), altitude ≤ 3000m.