Yokogawa DLM5000HD series high-definition oscilloscope

F: | Au:FAN | DA:2025-11-05 | 806 Br: | 🔊 点击朗读正文 ❚❚ | Share:

Collection restriction: When the record length in the ACQUIRE menu is set to the maximum value, the waveforms of LOGIC ports C (C0-C7) and D (D0-D7) cannot be collected.

Threshold and noise: The 701989 logic probe can set the threshold type (All/Each), and selecting the preset threshold will automatically configure it. After changing it, the preset will become "Userdef"; The noise suppression function is only available for probe 701989, and the level and noise suppression in the bit setting will be synchronized to the setting when the trigger source is Logic.

Vertical axis setting (SCALE/POSITION knob)

Analog signal: Press the CH1-CH8 keys to select the channel, adjust the vertical scale with the SCALE knob, and press the knob to enter FINE mode for high-precision adjustment; Adjust the vertical position with the POSITION knob, press it to set the position to 0 div. When adjusting, the corresponding channel information display area will temporarily display the relevant values, and there will be an out of range prompt.

Logic signal: Press the LOGIC key to control the display size of the logic signal on the SCALE knob, adjust the vertical position on the POSITION knob, and the LED will display the symbol color of the corresponding channel or logic signal.

All on/off (CH UTIL menu): It can collectively display or hide all analog signal channel waveforms, and logic signals do not have this collective control function.

Horizontal scale setting (TIME/DIV knob): Adjust the sensitivity of the horizontal scale, and temporarily display the time scale value and recording length in the upper right corner of the screen when controlling; ◄ POSITION ► Rotate the knob to adjust the horizontal position of the waveform, synchronously move the trigger position, press to set the trigger position to 50%, and adjust the waveform display in real-time during collection and operation.

(2) Trigger function

Trigger mode and hold time (MODE menu)

Trigger modes: including Auto (automatically updates waveform when triggering conditions are met or timeout occurs, with timeout of 100ms or the larger value of 10 time axis scales, and enables scrolling mode when the time axis exceeds the threshold), Auto Level (automatically sets the trigger level to the center value of the trigger source amplitude and triggers when timeout is not triggered), Normal (updates only when triggering conditions are met), N Single (stops and displays after collecting a specified number of signals), Single (presses the SINGLE button on the panel, collects waveform once and stops after triggering).

Holding time: Set through the Holdoff soft key to prevent accidental triggering, adjustable through jog shuttle or on-screen numeric keypad.

Collection Count: In N Count mode, the number of collected waveforms can be set, which can also be configured through jog shuttle or on-screen numeric keypad.

Trigger position and delay (◄ POSITION ►/DELAY key)

Trigger position: ◄ POSITION ► knob adjustment, temporary display of position value at the top of the screen during control, can also be set when no waveform is collected.

Trigger delay: After pressing the DELAY key to enable it, adjust the delay value with the ◄ POSITION ► knob, and the screen will temporarily display at the top. Press the DELAY key again to switch back to the trigger position setting; The delay cancellation function is enabled/disabled through the UTILITY menu system configuration. When enabled, the time measurement is based on the trigger position of 0s (no delay applied), and when disabled, the trigger point is 0s (delay applied), and the delay value is not affected by changes in the horizontal scale.

Various trigger settings (EDGE/ENHANCED/B TRIG menu)

Edge trigger (EDGE menu): Select the trigger source (CH1-CH8/Logic/EXT/LINE), set the trigger slope, high-frequency suppression, noise suppression, and trigger level. The settings for different sources (such as Logic requiring source position, EXT requiring probe attenuation and input range) may vary; LINE source only triggers rising edge.

Enhanced Trigger (ENHANCED Menu): Includes multiple types, such as multi edge trigger OR (select trigger source channel and slope, set trigger level for each channel, high-frequency/noise suppression), multi input mode trigger (set clock source, trigger source mode and combination conditions, time conditions, etc.), pulse width trigger (select source, polarity, high-frequency/noise suppression, trigger level, set time conditions and reference time, trigger when pulse width and reference time relationship are met), as well as rise/fall time trigger, runt signal trigger, timeout trigger, window trigger, etc. Some trigger types (such as bus signal trigger) are optional functions.

Combination trigger (B TRIG menu): Set the combination method of trigger A and trigger B (OFF/A Delay B/A to B (N)). In A Delay B mode, trigger A after it satisfies the specified time delay, and trigger B when it satisfies the specified time delay; In the A to B (N) mode, when A is triggered and B is triggered N times, it can be triggered, and both A and B can choose multiple types of edge triggering or enhanced triggering (such as bus signal triggering). However, when B is triggered through a window, the time limit is fixed at None, and the triggering condition for the mode without a clock source cannot be selected as True/False.

  • Basler D90 96801 100 PCB Card
  • Basler XR2002F Voltage Regulator (110 VAC, 48-480 Hz)
  • Basler SR8A-2B14B3A Regulator
  • Basler 9561500100 Module
  • Basler DECS-400 BE1-11 System
  • Basler DECS-100-B15 Excitation Control
  • Basler SCP 210 Frequency Controller
  • Basler SR4A-2B15B3A Static Voltage Regulator
  • Basler BE1-32R Power Relay
  • Basler PIA2400-17GM Power Interface Adapter
  • Basler MVC 232 Manual Voltage Control Module
  • Basler SSR 32-12 Static Voltage Regulator
  • Basler 5MW AVR Generator Voltage Regulator
  • Basler VR63-4B Voltage Regulator
  • Basler DECS-100-A05 AVR for Engine Generator
  • Basler DECS-100-B15 Automatic Voltage Regulator
  • Basler BE1-32R Directional Power Relay
  • Basler BE1-87B Differential Relay
  • Basler UFOV 260A Protective Module
  • Basler 9-2614-02-100 PCB Rev M
  • Basler DECS-100-B15 Digital AVR
  • Basler 9284900103 PS DECS-400N
  • Basler D4N3H1U Intertie Protection
  • Basler DECS-100-B15 A15 AVR
  • Basler KR4F Voltage Regulator
  • Basler BE26434 T14 Transformer
  • Basler SR8A-2B15B3A Regulator
  • Westinghouse 774B472A12 AR Relay
  • Basler DECS-100-B15 AVR
  • Basler XR2002F Regulator 110V
  • Basler SR125-E Static Regulator
  • Basler SSR 125-12 Regulator
  • Basler MOC2599 Motor Pot
  • Basler BE1-DFPR Feeder Relay
  • Basler CBS 305 Current Boost
  • Basler BE1-25 AutoSync
  • Basler MVC 300 Voltage Control
  • Basler BE3-25A AutoSync
  • Basler KR7FF Static Regulator
  • Basler 90-49000-100 Regulator
  • Basler 880 kVA Dry Type Transformer Specs
  • Basler Electric BE1-25 Sync-Check Relay Specs
  • Basler SSR 125-12 Voltage Regulator Specs
  • Basler Electric BE1-851 Overcurrent Relay Review
  • Basler Electric 149D930G02 Control Sub-Assembly
  • Basler Electric BE1-81O/UT Frequency Relay Specs
  • Basler Electric BE1-51/27C Overcurrent Relay
  • Basler Electric 149D956G02 Industrial Component
  • Basler Electric BE1-51A Overcurrent Relay Specs
  • Basler Electric BE1-40Q Loss of Excitation Relay
  • Basler DECS-200 Excitation Control System
  • Basler DECS-200 Voltage Regulator 56-277V AC / 125V DC
  • Basler BE1-87T Transformer Differential Relay
  • Basler RDP-110-S1 Protection Relay
  • Basler BE1-700V Digital Protective Relay
  • Basler BE1-951 Overcurrent Protection System
  • Basler DECS-300 Digital Excitation Control
  • Basler DECS-200 Digital Excitation Control
  • Basler DECS-200-1C Excitation Control System
  • Basler DECS-200-1L Digital Excitation Control
  • Basler Electric BE1-GPS Generator Protection System
  • Basler Electric DECS-200-1C Digital Excitation Controller
  • Basler Electric DECS125-15 Excitation Control with Power Module
  • Basler Electric BE1-87G Differential Relay
  • Basler Electric BE1-11 Protection System I5A3M2P2N0EA00
  • Basler Electric DECS-200-1C Excitation Control System
  • Basler Electric BE1-11g Generator Protection Relay
  • Basler Electric DECS 125-15-B2C1 V2.0.9 Excitation Control
  • Basler Electric BE1-81O/UT3ED1JA7N2F Frequency Relay
  • Basler Electric BE1-81O/UT3EE1YB7N1F Frequency Relay
  • Basler Electric DECS-200-1L Digital Excitation Control System
  • Basler DECS125-15-B2C1 Excitation Control
  • Basler 9507900205 SSR Retrofit Voltage Regulator
  • Basler BE2000E Digital Voltage Regulator
  • Basler BE1-GPS Generator Protection System
  • Basler DECS-250-CN1CN1N Digital Excitation Control
  • Basler DGC-2020 Genset Controller
  • Basler BE1-81O UT3ED1LA7N0F Frequency Relay (Variant)
  • Basler BE1-81O UT3EE1YA9S0F Frequency Relay (Variant)
  • Basler BE1-81O Over/Under Frequency Relay
  • Basler DECS125-15 Digital Excitation Control
  • Basler Electric BE1-951 Overcurrent Protection System
  • Basler Electric BE1-700V Digital Protective Relay
  • Basler Electric APR63-5 Automatic Voltage Regulator
  • Basler Electric BE1-851 Overcurrent Protection System
  • Basler Electric DECS-250-LN1SN1N Excitation Control
  • Basler Electric BE1-87T Transformer Differential Relay
  • Basler Electric DECS-200-1L Excitation Control System
  • Basler Electric 9310300100 DECS-300 Excitation Control
  • Basler Electric SSE-N 125-4.5KW Shunt Exciter Regulator
  • Basler Electric DGC-2020HD-5NS1DNSBA Genset Controller
  • Basler Electric BE1-81-O/UT3EE1JB7N1F Frequency Relay
  • Basler Electric BE1-81T1EE1WA0N1F Frequency Relay
  • Basler Electric BE1-25M1EA6PN5R1F Sync-Check Relay
  • Basler Electric BE1-GPS Generator Protection System
  • Basler Electric DECS-250-LN1SN1N Excitation Control Rev V
  • Basler Electric DECS-250-CN2CN1N Excitation Control
  • Basler Electric BE1-50/51B-207 Overcurrent Relay
  • Basler Electric DECS-300-C0N0 Excitation Control System
  • Basler Electric DECS-200 Digital Excitation Control System
  • Basler Electric DECS-250-LN1CN1N Excitation Unit
  • Basler Electric DECS-250 LN2SA1D Excitation Unit Specs
  • Basler Electric BE1-87T Transformer Relay Review
  • Basler Electric BE1-11 Protection System
  • Basler Electric BE1-GPS100-E4N1H1N Protection System
  • Allen-Bradley 442G-MABH-R Safety Module
  • Beckhoff CX1030-0111 PLC Assembly Profile
  • FANUC IC693CPU364 PLC Module
  • Orange Denmark Type 200816 220 PLC Specs
  • OMRON C200H-SNT31 Sysmac PLC Module
  • Allen Bradley 20AB022A3AYNANC0 PowerFlex 70
  • OMRON C200HW-PCU01 Position Control Unit
  • ABB AO845A-eA Analog Output Module
  • OMRON CJ1M-CPU22 CPU Unit
  • Allen Bradley 100-E265ED11 Contactor
  • Honeywell 51304511-100 Interface Module
  • SOLEXY BXF3S0101N0018 Gateway Module
  • OMRON CJ2H-CPU65 CPU Unit
  • Automation Direct GS2-45P0 AC Drive
  • M68-2000 2-Axis Motion CNC Controller
  • OMRON CJ1M-CPU11 V3.0 PLC CPU Unit
  • OMRON CJ1W-NC413 4-Axis Positioning Controller
  • OMRON 3G2A3-PRO16 Programming Console HMI
  • Siemens 3VT8440-2AA04-2GA2 Molded Case Circuit Breaker
  • Siemens 3RT5045 Contactor Series
  • OMRON C200HS-CPU01-E SYSMAC PLC Controller
  • OMRON C500-NC103-E Positioning Control Unit
  • OMRON CJ1W-TC001 Temperature Control Unit