Installation and Selection Guide
3.1 Flange selection
ISO-KF flange (DN 25/40): suitable for high vacuum applications (baking temperature ≤ 150 ° C), quick installation, using clamps and centering rings.
CF flange (DN 40 CF): suitable for ultra-high vacuum (UHV), can be baked up to 400 ° C, metal sealed.
Selection suggestion: If the system requires frequent baking or the maximum pressure is below 10 ⁻⁹ mbar, the CF flange version must be selected.
3.2 Installation location
The sensor should be installed in a location without direct process flow to avoid direct impact of particles and sputtered atoms on the measurement chamber.
It is recommended to install a baffle or bend between the sensor and the cavity to block contaminants.
Avoid installing sensors near the pump port to prevent contamination from pump oil vapor.
3.3 Magnetic field interference
The cold cathode sensor itself has a permanent magnet, therefore:
Ensure that there is no ferromagnetic material within 50 mm around the sensor during installation.
If there is a strong magnetic field in the system (such as sputtering cathode, electron beam evaporation source), magnetic shielding should be considered.
3.4 Electrical Connections
PR sensors typically have three electrodes:
Anode: connected to high voltage positive electrode (around+2 kV)
Cathode: connected to high voltage negative electrode (usually grounded)
Guard: Used to reduce leakage current (optional)
The controller needs to have:
High voltage output (0-5 kV, usually fixed at 2 kV)
Current measurement (pA level, logarithmic amplifier)
Ignition auxiliary circuit (improves starting reliability)
Use low-noise coaxial cables (such as RG-58) to connect sensors and controllers, with the shielding layer grounded at one end.

Ignition characteristics and improvement methods
Panning gauge may have difficulty igniting automatically when the pressure is below 10 ⁻⁴ mbar due to insufficient initial free electrons. The PR sensor uses a titanium cathode with a high secondary electron emission coefficient, but the following improvement measures still need to be noted:
Ensure sufficient initial pressure: When the system pressure is below 10 ⁻⁵ mbar, a small amount of gas (such as argon) can be introduced for a short period of time to 10 ⁻³~10 ⁻ ² mbar, and then lowered after the discharge is established.
High voltage scanning or ignition pulses using controllers: Many Leybold controllers (such as GRAPHIX, CENTER) provide automatic ignition functionality.
Clean cathode: The contaminated cathode surface will reduce the ignition success rate.
Cleaning and Maintenance: The Great Advantage of Removable Design
The long-term stability of a cold cathode sensor depends on the cleanliness of the measurement chamber. The oil vapor, sputtered atoms, and dust in the process will gradually deposit on the cathode and anode, resulting in:
Difficulty igniting
Current reading drift
Insulation resistance decreases (leakage current increases)
The easy disassembly and cleaning of PR sensors are their core selling points.
5.1 Cleaning steps
Release the system to atmospheric pressure and turn off the high pressure of the controller.
Remove the sensor (if the system is in vacuum, close the isolation valve first).
Use tools to open the sensor housing and remove the anode ring and cathode plate.
Cleaning method:
Mild pollution: Wipe with anhydrous alcohol or isopropanol, rinse with deionized water, and then dry.
Heavy pollution (such as metal plating): The surface of the cathode plate can be gently sanded with fine sandpaper (such as 600-1000 mesh) to remove the deposited layer.
Ultrasonic cleaning (in deionized water) is suitable for ceramic insulation components.
Check insulation resistance: Use a megohmmeter to measure the resistance between the anode and cathode, which should be>10 ¹² Ω.
Reassemble, pay attention to alignment.
Install on the system and test after vacuuming.
5.2 Replacement of cathode plate
When the cathode plate is worn or contaminated and cannot be cleaned, a replacement cathode plate (original parts) can be purchased. The manual states that "replaceable cathode plate" is a standard feature of PR sensors, which eliminates the need to replace the entire sensor and greatly reduces long-term usage costs.
5.3 Cleaning cycle recommendations
General laboratory application: once a year.
Coating or sputtering process: Every 3-6 months or based on reading drift.
Using a baffle: can extend the cleaning cycle by 2-3 times.
Troubleshooting
The following fault diagnosis is based on typical failure modes of Panning specifications.
6.1 No current output or display as zero (when pressure is below 10 ⁻⁴ mbar)
Possible cause inspection and resolution
Check the high voltage indicator light for controller high voltage not outputting; Measure high voltage output (pay attention to safety)
The sensor is not ignited (pressure is too low), introduce a small amount of gas to 10 ⁻ mbar, and observe whether the current jumps
Severe contamination of the cathode or anode leads to insulation degradation. Disassemble and clean, measure insulation resistance
Check coaxial cable connection for cable breakage or poor contact